137
Checking your meter
8
Did you use a suspended drop of control solution?
yes
Did you bend the test strip either before or while it was
drawing up control solution?
no
Did you wait for the flashing blood drop symbol to appear
before allowing the test strip to draw up control solution?
yes
Did you bend or move the test strip while the test was in
progress?
no
Did you compare the result with the concentration data spe-
cifically relating to the control solution you used?
yes
Is the concentration table on the test strip container from
which you removed the test strip?
yes
Is the measurement optics cover properly and firmly in place
on the meter (see Chapter 9 “Cleaning your meter”)?
yes
Did you measure within the correct temperature range
(+10 °C to +40 °C)?
yes
Summary of Contents for GO
Page 1: ...BLOOD GLUCOSE METER Instructions for use ...
Page 42: ...42 Entering settings Month Day ...
Page 66: ...66 Entering settings Entering alarm times Hours Minutes Alarm 2 ...
Page 187: ...187 Patents 18 18 Patents US 5 463 467 US 5 366 609 US 6 707 554 ...
Page 204: ...204 ...
Page 205: ......
Page 207: ...Warranty Card Name Street City post code Telephone No Date Signature ...