156
Measurement and storage conditions
Farbe: P 021 – Prüfmittelnummer: 201
O
Blood glucose tests can be performed at the limits of the permitted
range (b5 °C and +10 °C or b40 °C and +45 °C).
If the temperature is within one of these ranges, the thermometer
symbol flashes in the display.
Do not use results obtained at the limits of the permitted
temperature range as a basis for making therapeutic deci-
sions, as they could be incorrect. Incorrect results can cause
the wrong therapeutic decision to be taken and so lead to
serious adverse health effects.
O
Tests cannot be performed at temperatures below +5 °C and above
+45 °C.
In this case the following message is displayed: