188
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Result deletion
Result deletion algorithm. Automatic or
First in First out.
0: automatic
1: FIFO
1
N
Y
Result retention time (number of days
since a result was measured before it can
be deleted automatically if result deletion
algorithm is set to “automatic” and no
more memory is available).
1 – 1,000 d
30
N
Y
Result download required
0: disabled
1: enabled
0
N
Y
Electronic communication
Download warning
0 – 999 h
0: disabled
N
Y
Download Lockout
0 – 999 h
0: disabled
N
Y
Maximum number of list items transferred
in one POCT1-A message
1 – 500
75
N
Y
Application timeout (within this time the
application expects a response by the DMS
to any POCT1-A command)
30 – 3,600 s
60
N
Y
WLAN settings (general)
WLAN enabled
0: disabled
1: enabled
0
N
Y
Use DHCP
0: disabled
1: enabled
0
N
Y
IP (the meter‘s static IP-address if no DHCP
is used)
0.0.0.0 – 255.255.255.255 0.0.0.0
N
Y
Subnet Mask (subnet mask to be used by
the meter if static IP is used. Note: if DHCP
is used; the DHCP-subnet mask will be
used.)
0.0.0.0 – 255.255.255.255 0.0.0.0
N
Y
Use DNS
0: disabled
1: enabled
0
N
Y
DMS host (the host’s DNS name if DNS is
used)
0 – 60 characters
“”
N
Y
DMS IP address (the host’s IP address if no
DNS is used)
0.0.0.0 – 255.255.255.255 0.0.0.0
N
Y
DMS port (the port number to request
open socket from the DMS. Roche recom-
mends use of port numbers
1024.)
0 – 65,535
0
N
Y
DNS IP (IP address of the DNS-server if
DHCP is not used and DNS is configured)
0.0.0.0 – 255.255.255.255 0.0.0.0
N
Y
Gate IP (static gateway IP address)
0.0.0.0 – 255.255.255.255 0.0.0.0
N
Y
Subject/Attribute
Range
Default
Device
DMS
Summary of Contents for Inform II