Supplement for Observed Test Sequence
Observed Test Sequence (OTS)
The
Observed Test Sequence
(OTS) function allows an
observer (supervisor) to assess and record an operator’s
performance (e.g., for recertification purposes). The
observer monitors an operator during a test to check that
the test is being performed according to the recom-
mended procedures. He/she then evaluates the perfor-
mance and passes or fails the operator. This assessment
is saved together with the glucose test result and any
desired comments.
If so configured, the operator may receive a message
from the DMS when he or she tries to log in to the meter.
This message informs the operator that they have to
renew their certification. The contents of the message is
created on the DMS and may vary. The screen message
shown here is for illustration purposes only.
This function needs to be configured by the system
administrator and depends on available DMS settings.
Electronic configuration options will thus vary accord-
ing to the data management software utilized by your
institution.
Login Message
I-300:
Hello Dan Brown.
You have to perform an
OTS training before you
can use the meter.