92
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
The parameters for control solutions consist of four sepa-
rate values.
3
Use the keypad to enter the desired values one
after another:
– Minimum limit value for Level 1
– Maximum limit value for Level 1
– Minimum limit value for Level 2
– Maximum limit value for Level 2
4
Touch
to confirm each separate entry and
continue to the next entry.
1
2
3
4
5
6
7
8
9
0
Control L2 Max
353
mg/dL
12:48 pm
.
03/17/17
1
2
3
4
5
6
7
8
9
0
Control L2 Min
261
mg/dL
12:48 pm
.
03/17/17
1
2
3
4
5
6
7
8
9
0
Strip Expiration
05/30/17
12:48 pm
03/17/17