Maintenance and Care •
10
157
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
10.6 Unlocking a Download Lockout
When using a data management system for configura-
tion, a
Download Lockout
can be configured. This lockout
prevents a meter from being used for testing, if the meter
data have not been downloaded for a defined period of
time. If a download lockout occurs, the operator may use
a STAT test to temporarily override the lockout. If no STAT
test is configured or left, the meter cannot be used until
the stored data is downloaded via network (either wire-
less or by docking the meter).
In cases of e.g., network failures (which make a download
impossible), or other emergencies, it may be necessary to
unlock a meter without access to the data management
system. Since unlocking is a function of the
Setup Menu
,
it is restricted to persons who know the setup password.