232
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator’s Manual Version 6
Isolation Room Workflow
If configured, this feature allows you to reverse the order
of the steps “Entering or selecting the patient ID” and
“Confirming or selecting the test strip lot” to avoid con-
tamination issues. This workflow allows you to test a
patient in an isolation room without having to take the
test strip vial to the patient.
Test procedure for patients in an
isolation room
1
Confirm or select the test strip lot. For more details
2
Take the meter and the test strip to the isolation
room.
1
2
3
4
5
6
7
8
9
A-O
0
Patient ID
12:48 pm
03/17/17
Patient Test
Patient 123456789
Use Strip Lot
12:48 pm
123456?
03/17/17
Once a test strip has been removed from the vial,
the
blood sample must be applied within 3 minutes
.
Summary of Contents for Inform II
Page 3: ...08424705001 01 2017 04 USA Operator s Manual Version 6 0 A ACCU CCU CHEK HEK Inform II System ...
Page 245: ...245 Notes ...
Page 246: ...246 Notes ...
Page 247: ...247 ...