LANCING DEVICE

Instructions for Use

The ACCU‑CHEK Multiclix lancing device uses a lancet to obtain a drop of blood from a 

fingertip or other site. It is designed to make obtaining a drop of blood as comfortable as 

possible. The lancing device is loaded with a drum containing 6 sterile lancets. For 

virtually pain-free testing, we recommend that you advance to a new, sterile lancet after 

each test. You can adjust the lancet depth to your individual skin type by dialing 1 of the 

11 settings. There are 2 different caps: 1 is for a fingerstick, and the other is for other 

sites on the body. Refer to your meter owner’s booklet to determine what other sites are 

approved for your meter. 

The colored cap is for fingerstick testing only.

The clear cap is for testing from other sites of the body,  

not the fingertip.

Note: The ACCU‑CHEK Multiclix cap was not designed to be replaced after each 

use. The ACCU‑CHEK Multiclix lancing device is intended for use by a single person 

and is not suitable for testing multiple persons due to the risk of infections.

Inserting a Lancet Drum

You must first insert a lancet drum into the lancing device to get it ready for use.

A new drum has a blue ring that does not move.

A used drum has two visible red stripes and the blue ring can 

rotate on the white drum. You cannot insert a used drum into the 

lancing device.

1

  Remove the cap by pulling it 

straight off. It is not a twist cap. 

The easiest way to do this is to 

place your thumb on the cap 

grippers to the side of the notch 

as seen in the illustration.

2

  Before you insert a drum, be sure 

that the release button is not 

yellow. Insert a new drum, blue 

ring first, until it stops. It clicks 

into place.

3

  Slide the cap back on until it stops 

by aligning the notch on the cap 

with the notch on the device.

4

  You are now ready to use the first 

lancet. The lancet counter window 

on the plunger shows the number 

6, which means you have 6 new 

lancets remaining.

Performing a Fingerstick

Use a new lancet for each fingerstick. When you use a new lancet each time, you avoid 

infection and get a virtually pain-free test.

1

  Make sure the colored cap is on the lancing device. 

Do not use the clear cap for 

fingerstick testing.

2

  Adjust the depth by turning the 

comfort dial. The depth indicator 

shows the current depth selection. 

Higher numbers make the 

penetration deeper. If you have 

soft skin, start at a depth of 2. If 

your skin is calloused or thick, try 

a higher setting.

3

  Press the plunger in as far as it 

will go, like a pen.

4

  The release button turns yellow. Do not press the release button while pressing the 

plunger.

5

  Wash and dry your hands. Warm water helps to increase blood flow. If you use an 

alcohol wipe, be sure your finger is completely dry before continuing.

6

  Hold the lancing device firmly 

against the side of your fingertip. 

Remember that the opening where 

the lancet comes out is not in the 

center of the cap. Note: The side 

of your fingertip is less sensitive 

than the middle.

7

  Press the yellow release button.

8

  Gently squeeze your finger to 

assist the flow of blood. This helps 

you to get a blood drop.

9

  Apply the blood drop to your test strip.

0

  If you don’t get enough blood, try putting more pressure on your finger with the cap 

or switch to a higher setting. If you get too much blood, switch to a lower setting.

Loading the Next Lancet

1

  Twist the plunger one quarter turn 

forward (until it stops), then twist 

it back all the way. The lancet 

drum rotates to the next lancet.

2

  The number in the lancet counter 

window decreases by 1 and now 

shows the number 5.

3

  Each time you use a lancet and twist the plunger, the number showing in the lancet 

counter window decreases by 1. When the number 1 is visible in the lancet counter 

window, you have used 5 lancets. The sixth (and final) lancet is loaded and ready to 

be used.

Notes:

Only twist the plunger when the lancing device is not primed. You can see if the 

 

 

lancing device is primed by the color of the release button. A primed device has a 

yellow release button. When the lancing device is primed, the drum cannot rotate 

and twisting the plunger would damage the device. In rare cases, the lancet could 

protrude from the cap and you could injure yourself on it.
Only twist the plunger when at least “2” is displayed in the lancet counter window. 

 

 

If the number displayed in the lancet counter window is “1,” the last lancet is ready 

to be used. 

 

A built-in safety block stops you from twisting the plunger when the lancet counter 

window displays “1,” indicating there are no lancets left to load. 

 

Do not twist the plunger forcibly past the safety block. Damage to the lancing device 

may occur. This can override the protection against reusing used lancets. 
For safety reasons, once you advance to a new lancet, you cannot go back to a 

 

 

previously used lancet. Therefore, the plunger (and drum) cannot be twisted 

backwards. This built-in safety block prevents an infection risk, as used lancets 

cannot be reused.

Changing the Lancet Drum

When you have used the sixth and last lancet, you should change the drum.

1

  Take the colored cap off the lancing device by pulling it straight off. It is not a twist 

cap. The easiest way to do this is to place your thumb to the side of the notch as 

seen in the illustration.

2

  Hold the drum between your 

thumb and index finger and pull it 

straight out of the lancing device. 

Two red stripes are visible on the 

white drum, indicating that the 

drum has been used.

3

  Used lancet drums should always be disposed in accordance with local regulations.

4

  Insert a new drum, blue ring first, 

until it stops and clicks into place.

Notes:

Due to safety reasons, a used drum cannot be reinserted into the lancing device. 

 

 

 

A built-in safety block prevents this. This rules out a potential infection risk. 
In rare instances, if a used drum is dropped on a hard surface, the lancets may 

 

 

protrude slightly from the drum. If you drop a drum, always pick it up at the red-

striped end.

Getting Blood from a Site Other Than Your Fingertip

 

 

Capillary blood for blood sugar testing can be obtained not only from the fingertips but 

also from other sites (the palm, the forearm, the upper arm, the calf or the thigh), which 

is referred to as alternative site testing (AST). We recommend palm as the nonfingertip 

test site.
Use the clear ACCU-CHEK Multiclix AST cap together with the ACCU-CHEK Multiclix 

lancing device to collect blood from nonfingertip test sites. 
Your ACCU-CHEK blood glucose monitoring system is designed to test blood glucose also 

from sites other than your fingertip. Please refer to your test strip insert or your meter 

owner’s booklet for specific blood glucose and AST testing instructions.

IMPORTANT

 – There is a delay (lag time) in the delivery of blood glucose to nonfingertip 

sites when compared to fingertip. Therefore, there is a potential for a significant 

difference between fingertip and AST results. The length of delay (lag time) is strongly 

impacted by lifestyle, foods eaten or any other event having impact on blood glucose 

levels. It is important that people considering nonfingertip testing understand the 

following:

Talk to your doctor if you are considering testing on sites other than your fingertip.

 

 

AST is not recommended during periods of rapid decreases or increases in blood 

 

 

glucose levels.
AST is not recommended for people with a history of recurrent hypoglycemia, who 

 

 

suspect their blood glucose is extremely low, or who are unaware of hypoglycemic 

events.
Never ignore symptoms of high or low blood sugar.

 

 

Do not change your treatment based on one result.

 

 

If the AST reading does not match how you feel, perform a fingertip test to confirm 

 

 

your results. If your blood glucose still does not reflect the way you feel, contact 

your doctor. For more information on AST please contact the ACCU-CHEK Customer 

Care Service C

enter at 1-800-858-8072.

Do not test blood from the alternative sites:

U

 

 

p to 2 hours following a meal, when blood glucose values can change quickly

A

 

 

fter exercise

I

 

 

f you have a temperature or if a medical condition is limiting your mobility, e.g. after 

breaking a leg or being confined to bed
If you suspect that your blood glucose is extremely low (hypoglycemia)

 

 

D

 

 

uring peak action time of short-acting insulin (up to approximately 4 hours after 

the injection) or rapid-acting insulin analogues (up to approximately 2 hours after 

the injection)

1

  Remove the colored cap from the lancing device by pulling it straight off.

2

  Put the clear cap on by lining up the notched sections on the cap and the lancing 

device.

3

  Adjust the depth setting dial to the highest setting (5.5).

4

  Press the plunger in as far as it will go, like a pen.

5

  The release button turns yellow. Do not press the release button while pressing the 

plunger.

6

  Wash the nonfingertip site with warm water and dry it thoroughly.

7

  Press and hold the lancing device 

FIRMLY

 against the fleshy area on the chosen 

AST site. Press the lancing device up and down in a slow, pumping motion to assist 

the flow of blood. 

8

  Keep steady pressure on the area and press the yellow release button. Don’t be 

afraid to apply pressure to the area with the lancing device or with the clear AST 

cap alone to assist the flow of blood. Continue to press the lancing device or the 

clear cap against the site for a few seconds to allow blood to come to the surface.

9

  Perform a blood sugar test with your meter as instructed in your meter owner’s 

booklet.

Don’t be afraid to apply strong pressure to the area with the lancing device or with the 

clear AST cap alone to get a good blood drop.

Having Trouble Testing from Sites Other Than Your Fingertip?

If you experience difficulty obtaining a sufficient drop of blood while testing from a site 

other your fingertip, please contact our ACCU-CHEK Customer Care 

Service Center at 

1-800-858-8072 for assistance. We will guide you through the testing procedure step by 

step. If you continue to have difficulty, we can also offer you, free of charge, our 

ACCU-CHEK Softclix lancing device. Due to the ACCU-CHEK Softclix lancing device using 

a thicker diameter needle than the ACCU-CHEK Multiclix lancing device, you may have 

better success with the ACCU-CHEK Softclix lancing device when trying to test from sites 

other than your fingertip.

Caring for Your Lancing Device

Your lancing device requires very little care.

 

 

Keep it away from extreme temperatures.

 

 

Clean the outside of your lancing device regularly and always if any blood gets on it 

 

 

with 70% isopropyl (rubbing) alcohol to keep it in good working condition. In 

addition, thoroughly wipe the inside of the cap using a cotton ball lightly moistened 

with the 70% isopropyl (rubbing) alcohol. After cleaning, allow the lancing device 

time to dry thoroughly. Never dip the lancing device or the caps (colored cap and 

clear AST cap) in cleaning liquid. Doing so may damage the internal parts of the 

lancing device so affecting its operation.
Do not dip the lancing device in liquid.

 

 

Do not use bleach to clean the lancing device or caps.

 

 

Assistance Available

If you need assistance, please contact the ACCU-CHEK Customer Care 

Service Center 

toll-free, 24 hours a day, 7 days a week. We offer assistance in Spanish and many other 

languages.
1-800-858-8072 phone
1-800-858-8074 TDD/TYY (for hearing impaired)

One (1) Year Warranty

Roche warrants the ACCU-CHEK Multiclix lancing device against defects in materials and 

workmanship for a period of one (1) year from the date of purchase. The liability of 

Roche under this warranty is limited to the repair or replacement, at Roche’s option, of 

all parts or materials proven to be defective. This warranty does not cover repair or 

replacement of ACCU-CHEK Multiclix lancing devices which have been subjected to 

alteration, misuse, tampering, or abuse.
THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, WHETHER EXPRESS, 

IMPLIED, OR STATUTORY. ROCHE DISCLAIMS ANY AND ALL IMPLIED WARRANTIES, 

INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A 

PARTICULAR PURPOSE.
Roche’s entire liability with respect to the ACCU-CHEK Multiclix lancing device shall be 

the repair or replacement of nonconforming devices, at Roche’s option. In no case shall 

Roche be liable for incidental, indirect, special, or consequential damages. This warranty 

gives you specific legal rights, and you may have other rights which vary from state to 

state so the above limitation and exclusion may not apply to you. All inquiries about the 

ACCU-CHEK Multiclix lancing device under the above warranty or service policy must be 

directed to ACCU‑CHEK Customer Care Service C

enter by calling 1-800-858-8072. You 

will be informed of the procedure necessary for replacement. Please DO NOT return any 

device to Roche without authorization.

IMPORTANT

 – Read carefully before use. Caution: A restricted license limits use of the 

ACCU-CHEK Multiclix lancing device in the United States. The ACCU-CHEK Multiclix 

lancing device and its use are protected by U.S. Patent Nos. Re.35,803; 6,419,661; 

7,077,828; 7,223,276; 7,322,998. A license to use the ACCU-CHEK Multiclix lancing 

device is only granted when ACCU-CHEK Multiclix lancets are used with the ACCU-CHEK 

Multiclix lancing device.

ACCU‑CHEK Multiclix lancets are high precision components that are produced to the 

close tolerances required for satisfactory operation with the ACCU-CHEK Multiclix lancing 

device. Use of other lancets with the ACCU-CHEK Multiclix lancing device may prevent or 

impair proper function of the ACCU-CHEK Multiclix lancing device.

Using the ACCU‑CHEK Multiclix lancing device indicates your acceptance of the restricted 

license to use the ACCU-CHEK Multiclix lancing device only with ACCU-CHEK Multiclix 

lancets. Further, if you have purchased an ACCU-CHEK Multiclix lancing device that 

includes this restricted license, then this restricted license applies regardless of any 

additional offers found in ACCU-CHEK Multiclix lancing device packages. If you do not 

agree to the terms and conditions of the restricted license, you may return, at the place 

of purchase, the unused ACCU-CHEK Multiclix lancing device for a full refund. If you have 

any questions, please call the ACCU‑CHEK Customer Care Service Center at  

1-800-858-8072.
Except where prohibited by statute, all warranties covering the ACCU-CHEK 

Multiclix lancing device are voided by use of the ACCU-CHEK Multiclix lancing device with 

any lancets other than ACCU-CHEK Multiclix lancets.

Visit our 

Web site at www.accu-chek.com.

This product is protected by the following US Patents: 6,419,661; 7,077,828; 

7,223,276; 7,322,998; Re. 35,803.

 

This product fulfils the requirements of Directive 93/42/EEC on medical devices.

ACCU-CHEK and ACCU-CHEK MULTICLIX are trademarks of Roche.

Roche Diagnostics GmbH

 

Sandhofer Strasse 116 

 

68305 Mannheim

 

www.accu-chek.com

Manufactured in Germany for: 

 

Roche 

 

9115 Hague Road, Indianapolis, IN 46256, USA

©2010 Roche.

Cap

Pin‑hole  
opening

Used lancet drum 

with red stripe

Comfort dial for 
depth selection 

Release button

New lancet 

 

drum 

Depth 

 

indicator

Lancet counter window 

showing number of 

available lancets

Plunger for  

priming and loading  

next lancet

00000919 – 4 All-inhouse – 525x297 mm; 5-spaltig (breite) x 8-spaltig (höhe)

DE 05509068001(02) US 05509068002

V1/R1 (black) – 2010-06

0919 – 4all 3773 – (2010) Frau Heiland

pi_05509068001_02_En-US_525x297.indd   1

18.08.2010   10:23:13

Reviews: