EXC-100 MICROSCOPE SERIES
73 Mall Drive, Commack, NY 11725 • 631
-864-
1000 •
www.accu-scope.com
8
DETAILED ASSEMBLY
Fig. 3
Installing the Objectives
1. Rotate the coarse focusing knob
①
to lower the stage to a suitable
position (see Fig.3).
2. Install the objectives into the nosepiece
②
from the lowest
magnification to the highest in clockwise direction.
★
When operating, first use the low magnification objective (4X or 10X)
to search for specimen and focus, and then rotate in the high
magnification objective to observe.
★
When replacing the objective, rotate the objective nosepiece until it
sounds “ka-da”, to make sure the desired objective is in the center of
optical path.
Fig. 4
Installing the Eyepiece(s)
1. Remove the cover from the eyepiece tube
①
.
2. Insert the eyepiece
②
into the eyepiece tube, until it is completely
inserted.
3. Tighten eyepiece with M2.5 inner hexagon lock-screw
③
(see Fig.4).
Fig. 5
Connecting the Power Cord
★
To avoid damage to the power cord, to not use strong force when the
power cord is bent or twisted.
1.
Make sure the light adjustment knob is at “O”(OFF) before connecting
the power cord.
2. Fully insert the connector
①
of power cord into the power socket
②
,
and make sure it connects well (see Fig.5).
3. Fully insert the other connector into the socket of power supply, and
make sure it connects well.