EXC-120 MICROSCOPE SERIES
ACCU-SCOPE
®
8
DETAILED ASSEMBLY
(continued)
Fig. 5
Installing the Objectives
(Fig. 5 & 6)
Rotate the coarse focusing knob
①
counterclockwise to
lower the stage to its lowest position.
Install each objective
②
into the nosepiece
③
from the
lowest magnification to the highest in a clockwise direction
beginning with the first empty objective receptacle in front.
Install each objective by using two hands to position and
gently screw in the objective in a clockwise direction into
the threads of the nosepiece receptacle.
NOTE:
Never force any objectives onto the threads of the
nosepiece, and do not over-tighten.
Fig. 6
Fig. 7
Installing the Filters
(Fig. 7)
To install a filter, swing out the filter holder
①
located
below the condenser
②
by sliding it to the left using the
vertical handle
③
on the filter holder.
Place the filter
④
into the top of the holder.
Swing the filter holder
③
back into place under the
condenser.
①
②
③
①
②
③
④