EXC-120 MICROSCOPE SERIES
ACCU-SCOPE
®
11
ADJUSTMENT & OPERATION
(continued)
Fig. 13
Focusing -
(Fig. 13)
Set the left diopter collar
①
to “0”.
Using your left eye only and the 10X objective, focus your
specimen by adjusting the coarse focusing adjustment
knob
②
.
When the image is in view, refine the image to its sharpest
focus by turning the fine focusing adjustment knob
③
.
Rotate the diopter collar
①
to obtain the sharpest focus.
The diopter range is + or
– 5. The number aligned to the
white dot on the viewing head is the diopter in use.
NOTE:
do not counter rotate the focusing knob as this will
damage to the focusing system
.
Fig. 14
Adjusting the Condenser -
(Fig. 14)
The condenser was installed and centered in our factory by
our technicians prior to delivery. If the condenser needs to
be re-installed or adjusted in the future, follow the
instructions on page 13.
To adjust the condenser, turn the condenser focusing
knob
①
to raise or lower the condenser
②
.
Raise the condenser when using the highest magnification
objective, and lower it when using the low magnification
objective.
Focus the specimen with the 10x objective.
Adjust the condenser focusing knob
①
to get a clear
image of the field iris diaphragm.
①
③
②
①
②