![background image](http://html1.mh-extra.com/html/accu-scope/exc-500/exc-500_user-manual_3732920015.webp)
EXC-500 MICROSCOPE SERIES
ACCU-SCOPE
®
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
15
ADJUSTMENT & OPERATION
(continued)
“ECO” (Economy) Setting -
(Fig. 25)
The EXC-500 Microscope is equipped with an auto shut-off
feature, or “ECO” mode, to preserve the life of the bulb.
ECO mode activates when the microscope is unattended for
fifteen (15) minutes.
The ECO sensor
①
is located on the front of the microscope
base. While in ECO mode the red light of the ECO sensor will
flash once every three (3) seconds.
To reactivate the illumination, press the ECO button
②
once.
To
DISABLE
the ECO feature, PRESS and HOLD the ECO
button for three (3) seconds.
To
REACTIVATE
the ECO feature, PRESS and HOLD the ECO
button for three (3) seconds.
Replacing the Fuses -
(Fig. 26 & 27)
The fuse holder drawer is located underneath the AC power
cord receptacle on the back of the microscope.
To replace the fuses, turn the power off and unplug the
microscope from the wall outlet and then from the power cord
receptacle on the back of the microscope.
Insert a small flat head (-) screwdriver into the top of the fuse
holder drawer
①
and carefully slide it open to expose the two
fuses.
(Fig. 27 shows the view looking up from the bottom of the
microscope with the fuse holder drawer open.)
Being careful not to break the glass of either of the fuses, gently
insert the screwdriver up through the bottom openings of the
drawer as shown and push each fuse up to remove.
Replace the fuses with 250v T2.5A fuses, CAT# 500-3277
Slide the fuse holder drawer back in place.
LED BULB:
3w S-LED
INPUT:
AV 100 ~ 240v 50/60Hz
FUSE:
250v T2.5A
Fig. 25
①
②
Fig. 26