10
Purpose
With this application program you can add up successive weight values exerding capacity of balance/scale.
Example:
Totalizing weight values
Step
Key (or instruction)
Display
1. Select application program
(ZERO) > 2 sec
2. Select Totalizing
(F) repeatedly
3. Confirm setting
(ENTER)
Symbol “ “ on the display: application is active
4. If necessary: zero the balance/scale
(ZERO)
5. Place sample on balance/scale (in this example, 380 g)
6. Store value in memory. Total weight is displayed
(ENTER)
steadily; symbol
flashes.
7. Remove sample from balance/scale
8. Place the next sample on the balance/scale
(in this example, 575 g)
9. Store value in memory. Totalized stored weight
(ENTER)
is displayed; symbol flashes.
Note: symbol remains on indicating stored
value in memory until cleared
10. To view the current weight for 2 seconds
(F)
(if a printer is connected, a printout is generated)
11. Clear totalizing memory
(F) > 2 sec
(if a printer is connected, total is printed)
12. End Totalizing
(ENTER) > 2 sec
13. Reactivate Totalizing (if no other
(F)
application program has been selected)
(if a printer is connected, total is printed).
14. Repeat procedure starting from Step 6.
Totalizing