Specifications

Model

VIC-303

VIC-123

VIC-4MG VIC-2MG

VIC-612 VIC-412 VIC-212

Weighing capacity

g

300

120

410

210

610

410

210

Readability

g

0.001

0.001

0.005

0.005

0.01

0.01

0.01

Tare range (subtractive)

g

300

120

410

210

610

410

210

Linearity

±g 0.004

0.003

0.01

0.01

0.03

0.03

0.035

Operating temperature range

10°C to 30°C (273°K to 303°K; 50°F to 86°F)

Stabilization time (average)

s

2.5

2.5

2.5

2.5

2

2

2

Adaptation to ambient conditions

By selection of 1 of 4 optimized filter levels; display update: 0.1–0.8 (depends on filter level selected)

Calibration weight

g

200 (F1)

100 (F1)

200 (F1)

200 (F2)

200 (F2)

200 (F2)

200 (M1)

Net weight, approx. 

kg

1.3

1.2

1.3

1.3

1.35

1.35

1.2

Pan size

mm

97 

d

97 

d

97 

d

97 

d

142x130

142x130

97 

d

Power source/voltage/frequency

AC adapter, 230 V or 115 V, +15% to – 20%, 48–60 Hz

Power consumption (average)

W

1

1

1

1

0.75

0.75

0.75

Hours of operation w/ 9V battery:
– Alkaline (approx.)

h

11

11

14

– Rechargeable, fully ch., 

(NiMH), avg.

h

2.5

2.5

4

Model

VIC-5101 VIC-3101 VIC-1501 VIC-711 VIC-511 VIC-10KG VIC-6KG VIC-4KG

Weighing capacity

g

5100

3100

1500

710

510

10100

6100

4100

Readability

g

0.1

0.1

0.1

0.1

0.1

1

1

1

Tare range (subtractive)

g

5100

3100

1500

710

510

10100

6100

4100

Linearity

±g 0.2

0.2

0.2

0.2

0.2

2

2

1

Operating temperature range

10°C to 30°C (273°K to 303°K; 50°F to 86°F)

Stabilization time (average)

s

2

2

1.5

1.5

1.5

1.5

1.5

1.5

Adaptation to ambient conditions

By selection of 1 of 4 optimized filter levels; display update: 0.1–0.8 (depends on filter level selected)

Calibration weight

kg

5 (F2)

2 (F2)

1 (M1)

0,2 (M2) 0,2 (M2) 5 (M1)

5 (M2)

2 (M2)

Net weight, approx. 

kg

1.1

1.1

1.1

1.25

1.25

1.1

1.1

1.1

Pan size

mm

142x130

Power source/voltage/frequency

AC adapter, 230 V or 115 V, +15% to – 20%, 48–60 Hz

Power consumption (average)

W

1

1

0.75

0.75

0.75

0.75

0.75

0.75

Hours of operation w/ 9V battery:
– Alkaline (approx.)

h

11

11

14

14

14

14

14

14

– Rechargeable, fully ch., 

(NiMH), avg.

h

2.5

2.5

4

4

4

4

4

4

Overview

Visit www.balances.com for the best prices on Acculab scales

www.balances.com is the original deep discount distributor

Summary of Contents for 98648-013-61

Page 1: ...ULAB VICON Electronic Precision Scales Balances Elektronische Pr zisionswaagen Balances lectroniques d analyse et de pr cision Bilance elettroniche di precisione Balanzas electr nicas de precisi n Vis...

Page 2: ...autions Safety Information To prevent damage to the equip ment please read these operating instructions carefully before using your balance scale Do not use this equipment in hazardous areas Make sure...

Page 3: ...top of one another at a time Installation Choose a location that is not subject to the following nega tive influences Heat heater or direct sunlight Drafts from open windows and doors Extreme vibrati...

Page 4: ...ve the compartment cover Install the battery in the compart ment Make sure the polarity is correct Close the battery compartment slide the cover into position until it snaps into place Do not throw aw...

Page 5: ...e Keys ON OFF On off key switches the balance scale on and off or switches it to the standby mode Battery operation on turns backlight on off ZERO Zeros the balance press and hold 2 seconds opens the...

Page 6: ...version factor Display 1 user Grams 1 00000000000 o 2 grams factory setting Grams 1 00000000000 g 4 carat Carats 5 00000000000 o 5 Pound Pounds 0 00220462260 lb 6 ounce Ounces 0 03527396200 oz 7 tryo...

Page 7: ...uantity 20 on the balance scale 7 Select reference sample quantity in increments of 1 1 2 3 99 or F repeatedly briefly or in increments of 10 10 20 30 100 F 2 sec 8 Confirm selected reference sample q...

Page 8: ...6 Place the reference weight for 100 on the balance scale 7 If desired change the number of decimal places displayed 100 0 100 00 or 100 factory setting F repeatedly 8 Confirm selected no of decimal p...

Page 9: ...alance scale ZERO 5 Place oversized sample on balance scale 6 Start application program F Symbol flashes on the display the weight value is locked 7 Unload the balance scale the weight value remains d...

Page 10: ...ce scale 8 Place the next sample on the balance scale in this example 575 g 9 Store value in memory Totalized stored weight ENTER is displayed symbol flashes Note symbol remains on indicating stored v...

Page 11: ...display airval ENTER 7 Determine weight of the sample in air place sample on the balance scale 8 Store weight in air value ENTER 9 Remove sample from balance scale 10 Determine weight of sample in liq...

Page 12: ...libration weight value and ENTER start calibration span adjustment After the zero point is stored the required calibration weight flashes on the display 6 Place the required calibration weight on the...

Page 13: ...1 3 Stability range 1 3 1 1 4 digit 1 3 2 1 2 digit 1 3 3 1 digit 1 3 4 2 digits 1 3 5 4 digits 1 5 Calib adjust linearization 1 5 1 Calibration adjustment Function of the CAL key 1 5 2 Linearization...

Page 14: ...aded Unload the balance scale appErr Weight is too light or Increase the weight on the balance scale there is no sample on the balance scale with application in use PrtErr Data interface for Contact t...

Page 15: ...eable fully ch NiMH avg h 2 5 2 5 4 Model VIC 5101 VIC 3101 VIC 1501 VIC 711 VIC 511 VIC 10KG VIC 6KG VIC 4KG Weighing capacity g 5100 3100 1500 710 510 10100 6100 4100 Readability g 0 1 0 1 0 1 0 1 0...

Page 16: ...Standards Limitation of emissions In accordance with product standard EN 61326 1 Class B residential area Defined immunity to interference in accordance with product standard EN 61326 1 minimum test r...