Basic Weighing Function
Features
– Zeroing the balance
You can zero the balance/scale within the entire weighing
range, up to the maximum capacity.
Preparation
§
Switch on the balance/scale: press the (ON/OFF) key
§
If necessary, zero the balance/scale: press the (ZERO) key
$
If necessary, change the configuration settings:
see the chapter entitled “Configuration”
$
If desired, load the factory settings:
see the chapter entitled “Configuration,” menu item 9.– 1
Additional functions:
$
Switching off the balance/scale: press the (ON/OFF) key
$
Battery operation: automatic shut-off after 2,5 or 10 min-
utes; see chapter on configuration. Example: 2 minutes.
If the weight readout remains unchanged and no keys are
pressed for at least two minutes, the battery symbol
“
“ starts flashing. After another 2 seconds, the
balance shuts off automatically, unless a key is pressed.
Description of the Keys
(ON/OFF)
On/off key: switches the balance/scale on and
off or switches it to the standby mode.
Battery operation: on; turns backlight on; off
(ZERO)
Zeros the balance; press and hold 2 seconds:
opens the application menu
(CAL)
Starts calibration/adjustment
(F)
Starts an application program;
Scrolling in application menu, configration
menu and calibration menu
(ENTER)
Confirms the selected setting;
Exits application, configuration & calibration
menu if key is pressed and held for more than
2 seconds.
(PRINT)
Generates a printout or data output
Operation
Example:
Determine weight of sample
Step
Key (or instruction)
Display
1. Switch on the balance/scale
(ON/OFF)
Self-test runs
Display: Software version
r 31.01
2. Open the in-use cover and leave open
while weighing
3. Place container on the balance/scale
52.0
g
(in this example, 52 g)
4
. Zero the balance/scale
(ZERO)
0.0
g
5. Place sample in container on balance/scale
150.2
g
(in this example, 150,2 g).