9
Purpose
“Holds” the displayed value; also, the display will be locked for 5 seconds after removing the sample from the pan.
Example:
Determine weight of oversized sample
Step
Key (or instruction)
Display
1. Select application program
(ZERO) > 2 sec
2. Select Hold Display
(F) repeatedly
3. Confirm setting
(ENTER)
Symbol “ “ on the display: application is active
4. If necessary: zero the balance/scale
(ZERO)
5. Place oversized sample on balance/scale
6. Start application program
(F)
Symbol “ “ flashes on the display:
the weight value is locked
7. Unload the balance/scale: the weight value
remains displayed for a further 5 seconds; or
8. Zero the balance/scale
(ZERO)
9. End the Display Hold application
(ENTER) > 2 sec
10. Reactivate Display Hold (if no other
(F)
application program has been selected)
11. Repeat procedure starting from Step 5.
“Hold” Display