14
Error codes are shown on the main display for approx. 2 seconds. The program then returns automatically to the previous mode.
Display/Problem
Cause
Solution
No segments appear
No power available
Check the power supply
on the display
The AC adapter is not plugged in
Plug in the AC adapter
Battery is drained
Replace battery; recharge battery using
external charger
oL
The load exceeds the balance/scale
Unload the balance/scale
capacity
UL
Weighing pan not in place
Place the weighing pan
on the balance/scale
Something is touching
Move the object that is
the weighing pan
touching the weighing pan
disErr
Display overflow:
Reduce load on the balance/scale
Value cannot be shown
on the display
calErr
Calibration parameter not met; e.g.:
Calibrate only when zero is displayed
– Balance/scale not zeroed
Press (ZERO) to tare the balance/scale
– Balance/scale is loaded
Unload the balance/scale
appErr
Weight is too light or
Increase the weight on the balance/scale
there is no sample on the balance/scale
with application in use
PrtErr
Data interface for
Contact the Acculab
printing is blocked
customer service center
Max. weighing capacity is less
The balance/scale was switched on
Place the weighing pan on the
than indicated under “Specifications”
without the weighing pan in place
balance/scale and press (ON/OFF)
The weight readout is obviously wrong
The balance/scale was not calibrated/
Calibrate/adjust the balance/scale
adjusted before weighing
Balance/scale not zeroed
Zero the balance/scale
If any other errors occur, contact your local Acculab customer service center.
Error Codes