15
Specifications
Model
VIC-303
VIC-123
VIC-4MG VIC-2MG
VIC-612 VIC-412 VIC-212
Weighing capacity
g
300
120
410
210
610
410
210
Readability
g
0.001
0.001
0.005
0.005
0.01
0.01
0.01
Tare range (subtractive)
g
300
120
410
210
610
410
210
Linearity
≤
±g 0.004
0.003
0.01
0.01
0.03
0.03
0.035
Operating temperature range
10°C to 30°C (273°K to 303°K; 50°F to 86°F)
Stabilization time (average)
s
2.5
2.5
2.5
2.5
2
2
2
Adaptation to ambient conditions
By selection of 1 of 4 optimized filter levels; display update: 0.1–0.8 (depends on filter level selected)
Calibration weight
g
200 (F1)
100 (F1)
200 (F1)
200 (F2)
200 (F2)
200 (F2)
200 (M1)
Net weight, approx.
kg
1.3
1.2
1.3
1.3
1.35
1.35
1.2
Pan size
mm
97
d
97
d
97
d
97
d
142x130
142x130
97
d
Power source/voltage/frequency
AC adapter, 230 V or 115 V, +15% to – 20%, 48–60 Hz
Power consumption (average)
W
1
1
1
1
0.75
0.75
0.75
Hours of operation w/ 9V battery:
– Alkaline (approx.)
h
–
–
–
–
11
11
14
– Rechargeable, fully ch.,
(NiMH), avg.
h
–
–
–
–
2.5
2.5
4
Model
VIC-5101 VIC-3101 VIC-1501 VIC-711 VIC-511 VIC-10KG VIC-6KG VIC-4KG
Weighing capacity
g
5100
3100
1500
710
510
10100
6100
4100
Readability
g
0.1
0.1
0.1
0.1
0.1
1
1
1
Tare range (subtractive)
g
5100
3100
1500
710
510
10100
6100
4100
Linearity
≤
±g 0.2
0.2
0.2
0.2
0.2
2
2
1
Operating temperature range
10°C to 30°C (273°K to 303°K; 50°F to 86°F)
Stabilization time (average)
s
2
2
1.5
1.5
1.5
1.5
1.5 1.5
Adaptation to ambient conditions
By selection of 1 of 4 optimized filter levels; display update: 0.1–0.8 (depends on filter level selected)
Calibration weight
kg
5 (F2)
2 (F2)
1 (M1)
0,2 (M2) 0,2 (M2) 5 (M1)
5 (M2)
2 (M2)
Net weight, approx.
kg
1.1
1.1
1.1
1.25
1.25
1.1
1.1
1.1
Pan size
mm
142x130
Power source/voltage/frequency
AC adapter, 230 V or 115 V, +15% to – 20%, 48–60 Hz
Power consumption (average)
W
1
1
0.75
0.75
0.75
0.75
0.75
0.75
Hours of operation w/ 9V battery:
– Alkaline (approx.)
h
11
11
14
14
14
14
14
14
– Rechargeable, fully ch.,
(NiMH), avg.
h
2.5
2.5
4
4
4
4
4
4
Overview