background image

4

Connecting the
Balance/Scale to AC Power

§

It is recommended to use only the
included AC adapter for optimal 
performance and safety.

§

Insert plug into the jack 
(located on back of scale)

§

Plug the AC adapter into an electrical
outlet

!

LISTED power supply 11 V–21 V
compliant with NEC Class 2 output.

Below-Balance/Scale 
Weighing

A port for a below-balance weighing
hanger is located on the bottom of
the balance/scale.

§

Open the cover plate on the bottom
of the balance/scale.

§

Attach the sample (e.g., using a sus-
pension wire) to the hanger.

$

Install a shield for protection against
drafts if necessary.

Installing the Battery

(not for models VIC-303, VIC-123,
VIC-4MG, VIC-2MG)

$

Batteries are not included with the
equipment supplied

!

Use only commercially available 9V
batteries. 

!

If you use a rechargeable battery,
recharge it with an external battery
charger.

§

Lay the balance/scale on its side

§

Open the battery compartment:
remove the compartment cover

§

Install the battery in the compart-
ment

$

Make sure the polarity is correct.

!

Close the battery compartment: slide
the cover into position until it snaps
into place

!

Do not throw away used batteries
with normal household waste. 
Rechargeable batteries contain toxic
materials and must be disposed of in
accordance with local waste disposal
regulations.

Leveling the Balance/Scale 

(only for models VIC-303, 
VIC-123, VIC-4MG, VIC-2MG)

§

Always level the balance/scale
again any time after it has been
moved to a different location.
Example: moving bubble from R to L

§

Turn the feet as shown in the dia-
gram until the air bubble is centered
within the circle of the level indicator.

> In most cases this will require several

adjustment steps.

Removing Weights for
Calibration/Span Adjustment

(only for models VIC-612, VIC-212,
VIC-711, VIC-511, VIC-303, 
VIC-123, VIC-4MG, VIC-2MG)

§

Grasp the tab to turn and remove
the weight compartment.

$

Follow instructions on page 12 for
calibration/span adjustment.

Summary of Contents for ACCULAB VICON VIC-123

Page 1: ...de d emploi Istruzioni per l uso Manual de instrucciones ACCULAB VICON Electronic Precision Scales Balances Elektronische Pr zisionswaagen Balances lectroniques d analyse et de pr cision Bilance elett...

Page 2: ...ies Options 16 C Marking 2 Contents Warnings and Safety Precautions Safety Information To prevent damage to the equip ment please read these operating instructions carefully before using your balance...

Page 3: ...adjustable feet Storage Do not stack more than 3 balances on top of one another at a time Installation Choose a location that is not subject to the following nega tive influences Heat heater or direct...

Page 4: ...lance scale on its side Open the battery compartment remove the compartment cover Install the battery in the compart ment Make sure the polarity is correct Close the battery compartment slide the cove...

Page 5: ...off automatically unless a key is pressed Description of the Keys ON OFF On off key switches the balance scale on and off or switches it to the standby mode Battery operation on turns backlight on of...

Page 6: ...e on balance scale 7 Toggle weight unit F Menu code Unit Conversion factor Display 1 user Grams 1 00000000000 o 2 grams factory setting Grams 1 00000000000 g 4 carat Carats 5 00000000000 o 5 Pound Pou...

Page 7: ...ale 5 Zero the balance scale ZERO 6 Place reference sample quantity 20 on the balance scale 7 Select reference sample quantity in increments of 1 1 2 3 99 or F repeatedly briefly or in increments of 1...

Page 8: ...mpty container on the balance scale 5 Tare the balance ZERO 6 Place the reference weight for 100 on the balance scale 7 If desired change the number of decimal places displayed 100 0 100 00 or 100 fac...

Page 9: ...n the display application is active 4 If necessary zero the balance scale ZERO 5 Place oversized sample on balance scale 6 Start application program F Symbol flashes on the display the weight value is...

Page 10: ...yed ENTER steadily symbol flashes 7 Remove sample from balance scale 8 Place the next sample on the balance scale in this example 575 g 9 Store value in memory Totalized stored weight ENTER is display...

Page 11: ...ance scale ZERO 5 Start application program F 6 Confirm the display airval ENTER 7 Determine weight of the sample in air place sample on the balance scale 8 Store weight in air value ENTER 9 Remove sa...

Page 12: ...different calibration weight value F repeatedly 5 Confirm calibration weight value and ENTER start calibration span adjustment After the zero point is stored the required calibration weight flashes on...

Page 13: ...ons 1 2 Application filter 1 2 1 Final readout 1 2 2 Filling 1 3 Stability range 1 3 1 1 4 digit 1 3 2 1 2 digit 1 3 3 1 digit 1 3 4 2 digits 1 3 5 4 digits 1 5 Calib adjust linearization 1 5 1 Calibr...

Page 14: ...play calErr Calibration parameter not met e g Calibrate only when zero is displayed Balance scale not zeroed Press ZERO to tare the balance scale Balance scale is loaded Unload the balance scale appEr...

Page 15: ...of operation w 9V battery Alkaline approx h 11 11 14 Rechargeable fully ch NiMH avg h 2 5 2 5 4 Model VIC 5101 VIC 3101 VIC 1501 VIC 711 VIC 511 VIC 10KG VIC 6KG VIC 4KG Weighing capacity g 5100 3100...

Page 16: ...request Accessories Options C Marking The balance scale complies with the following EC Directives and European Standards Council Directive 89 336 EEC Electromagnetic compatibility EMC Applicable Europ...

Reviews: