External Calibration
Parameters (changes in factory settings):
SETUP
-
Waage
-
CAL.Just.
-
Cal.Ext.
(menu code
1.1. 9. 1
)
The required calibration weight is configured at the factory (see “Specifications”)
Step
Key (or instruction)
Display
1. Unload and tare the balance
(TARE)
0.0
g
2. Start calibration.
(CAL)
Cal.Ext.
g
Once you store the zero point
-
5000.0
g
the required calibration weight
is prompted (flashing display)
3. Apply the prompted calibration
5000.0
g
weight (in this example: 5000 g)
Weight too light:
a minus sign “–” is shown
Weight too heavy:
a plus sign “+” is shown
The display stops flashing
as soon as the weight value is
within the defined limit.
4. Calibration/adjustment executed;
Cal.end
g
then the calibration weight
is displayed
+
5000.0
g
5. Remove the calibration weight
0.0
g
16