background image

Example: 

Determining residual weight in percent

Parameter settings: 

Applic.  Percent

(menu code 2. 4.)

Reference percentage:  

Ref 100

%

Step

Key (or instruction)

Display/Data output

1. Tare the balance

(TARE)

0.0

g

2. Place sample equal to 100% 

on the balance 
(in this example: 111.6 g)

3. Information:

Enter reference percentage

(MENU)*

ref     100

%

(Changing the reference: see the previous page)

4. Initialize the balance

(MENU)*

+

100.0

%

*

The current reference weight 

pRef        100 %

remains stored until a new 

Wxx%  +   111.6 g

reference is set or the power supply is interrupted

5. Remove sample (e.g., for drying)

6. Place unknown weight on balance 

+

94.9

%

*

(in this example: 105.9 g)

7. If desired, print percentage

(PRINT)

Prc   +    94.9 %

8. Toggle display between 

Repeatedly: 

+

105.9

g

*

weight and percentage

(MENU)*

+

94.9

%

*

9. Clear display of residual weight 

(CLEAR)

+

105.9

g

and reference percentage

10. If desired, print net residual weight

(PRINT)

N     +    05.9 g

* On some models, the keypad overlay shows the German word “MENUE“ for MENU.

30

Summary of Contents for Atilon

Page 1: ...98648 015 55 Operating Instructions ACCULAB ATILON Electronic Analytical and Precision Balances...

Page 2: ...20 Parameter Settings Overview 24 Input ID Number Date and Time 26 Application Programs 27 Counting 29 Weighing in Percent 31 Calculation 33 Animal Weighing Averaging 35 Net total Formulation 37 Total...

Page 3: ...connecting the cable check all pin assignments against the cabling diagrams and disconnect any lines that are assigned differently If there is visible damage to the equipment or power cord disconnect...

Page 4: ...models ATL 623 423 153 Round glass draft shield with cover Getting Started Installation Choose a location that is not subject to the following negative influences Heat heater or direct sunlight Drafts...

Page 5: ...tion the components listed below in the order given Place the lower lid on the balance with the raised edge facing upwards Pan support Weighing pan Glass draft ring Place the upper lid on the draft sh...

Page 6: ...Acclulab AC adapter with a wide input voltage range 100 to 240 V order no 6971966 and replaceable power cord 6900900 Europe 6900905 AUS 6900901 US CDN 6900902 ZA 6971945 UK 6971977 Argentina 6971973...

Page 7: ...e in which case the user will be required to correct the interference at his own expense If you have a Class B digital device please read and follow the FCC information given below However there is no...

Page 8: ...after this time will the instrument have reached the required operating temperature Using Verified Balances in Legal Metrology in the EU Make sure to allow the equipment to warm up for at least 24 hou...

Page 9: ...is not permitted in legal metrology Open cover plate on the bottom of the balance Important set the balance on its side to access the cover plate DO NOT turn the balance upside down Using the built i...

Page 10: ...front feet as shown in the diagram until the air bubble is centered within the circle of the level indicator In most cases this will require several adjustment steps On models with a rectangular weig...

Page 11: ...ity icon after you turn on the power A will be displayed until you press a key 12 Weight value displayed in selected weight unit Symbols Save settings and exit the operating menu One menu level higher...

Page 12: ...ruments in the EU The type approval certificate for verification applies only to non automatic weighing instruments For automatic operation with or without auxiliary measuring devices you must comply...

Page 13: ...32 30 1 Switch on the balance ON OFF 0 0 g Self test is performed followed by automatic initial tare function 2 Place container on weighing pan 11 5 g in this example 11 5 g 3 Tare the balance TARE 0...

Page 14: ...adjustment CAL Einh gramm Kilogr or Pound not for verified models To block calibration adjustment Select Cal Adj blocked in the menu Close the menu access switch on the back of the balance External C...

Page 15: ...built in motorized calibration weight is applied and removed automatically for internal calibration Step Key or instruction Display 1 Unload and tare the balance TARE 0 0 g 2 Start calibration CAL CA...

Page 16: ...CAL Cal Ext g Once you store the zero point 5000 0 g the required calibration weight is prompted flashing display 3 Apply the prompted calibration 5000 0 g weight in this example 5000 g Weight too lig...

Page 17: ...Scroll through menu items Press and hold ENTER One menu level lower ENTER Confirm menu item CLEAR Save settings and exit menu from any position Press and hold CLEAR Save settings and exit menu CLEAR...

Page 18: ...peatedly ENTER english o scrolls to the right 4 Change setting MENU Espanol Scroll until the desired setting is shown 5 Confirm the menu code ENTER Espanol o o indicates the active setting 6 Return to...

Page 19: ...ons Signal Acoustic signal beep 1 8 2 Keys Keypad 1 8 3 ext Keys Funktion external Keypad 1 8 4 On Mode Power on mode 1 8 5 Backlit Display backlighting 1 8 6 reset menu Factory settings 1 9 1 Applic...

Page 20: ...1 1 3 3 2 Digit 1 1 3 4 4 Digit 1 1 3 5 8 Digit1 1 1 3 6 Taring 1 W o Stb W o stability 1 1 5 1 Taring o W StAb After stability 1 1 5 2 AutoZer Off 1 1 6 1 Auto zero On 1 1 6 2 Wt Unit For list of un...

Page 21: ...1 6 1 1 Printing fct manual Man With W stability 1 6 1 2 automatic Aut W o Autom w o stability 1 6 1 3 Aut With Autom w stability 1 6 1 4 StopAut Stop Off Not possible 1 6 2 1 automatic printing On Us...

Page 22: ...Day month year 1 6 9 1 mmm dd yy Month day year 1 6 9 2 Extras Menu CanEdit Can change settings 1 8 1 1 Additional Rd Only Read only 1 8 1 2 functions Signal Off 1 8 2 1 Acoustic signal On 1 8 2 2 Key...

Page 23: ...5 1 1 Net total Component On 2 5 1 2 printout Total Comp Prt Off 2 6 1 1 Totalizing Component On 2 6 1 2 printout AnimalW Activty Calm fluct 2 of test obj 2 7 1 1 Animal Animal activity Active fluct...

Page 24: ...tem for setting the date 3 2 TIME Menu item for setting the time 3 3 Depending on the setting you select for the menu item SETUP PRNT OUT DATE the date will be displayed in the following format Format...

Page 25: ...R FEB 08 8 Change setting MENU FEB 10 hold down key to automatically change the digit s MENU hold 9 Toggle between ENTER or FEB 10 Day Month Year positions CLEAR 10 Save setting when you reach the YEA...

Page 26: ...xt or Codes Languag English factory setting 5 1 Deutsch German 5 2 Franc French 5 3 ital Italian 5 4 Espanol Spanish 5 5 pyccKV 1V 1 Russian 5 6 polski Polish 5 7 codes Menu shows codes not texts 5 8...

Page 27: ...matic o Factory setting Changing the Reference Sample Quantity Activate function Press the MENU key Select the desired reference sample quantity 1 to 100 In increments of 1 Press the MENU key briefly...

Page 28: ...100 MENU press and hold ref 20pcs 6 Confirm selected reference sample quantity ENTER 20pcs and start application The current reference nRef 20 pcs weight remains stored until a new reference wRef 1 0...

Page 29: ...ce 2 Dec Pl 2 decimal places 3 Dec Pl 3 decimal places o Factory setting Changing the Reference Percentage Activate function Press the MENU key Select the desired reference 1 to 100 In increments of 1...

Page 30: ...ize the balance MENU 100 0 The current reference weight pRef 100 remains stored until a new Wxx 111 6 g reference is set or the power supply is interrupted 5 Remove sample e g for drying 6 Place unkno...

Page 31: ...cs Decimal places None No Decimal places o 1Dec Pl 1 Decimal places 2Dec Pl 2 Decimal places 3Dec Pl 3 Decimal places o Factory setting Setting the Factor or Divisor Activate function Press the MENU k...

Page 32: ...0 06237 ENTER Position the decimal point 5 MENU __ 00000 2 ENTER Enter numerals Repeatedly __ 06000 MENU or press and hold ENTER etc __ 06237 4 Store the divisor and initialize the balance The curren...

Page 33: ...ram AnimalW Animal weighing activty Animal activity Calm Stable readout o Active Unstable V Active Very unstable readout Start manual o Autom Automatic o Factory setting Changing the Number of Subweig...

Page 34: ...ts and start ENTER 0 0 g automatic animal weighing The number of measurements remains stored in battery backed memory until the setting is changed 6 Place first animal in bowl 888 The balance delays t...

Page 35: ...Features Weigh up to 99 components from 0 to a defined total component weight Store component weights printout shows Comp xx with display zeroed automatically after value is stored and automatic print...

Page 36: ...4 Store component data ENTER 0 0 g NET Comp 1 120 5 g 5 Add next component 70 5 g NET 6 Store component data ENTER 0 0 g NET Comp 2 70 5 g 7 Weigh in further components Repeat steps as desired 5 and...

Page 37: ...Off o On o Factory setting Features Totalizing memory for up to 99 values Store component weights printout shows Comp xx with automatic printout Toggle display between the current individual weight v...

Page 38: ...le 380 g 3 Store value in memory ENTER 380 0 g Comp 1 380 0 g 4 Remove sample 0 0 g 5 Place the next sample on the balance in this example 575 g 575 0 g 6 Store value in memory ENTER 955 0 g 575 0 g C...

Page 39: ...d in the example below are not included with the balance Preparation Select the Density Determination application in the menu see Configuration Set the following parameters Applic Application program...

Page 40: ...Confirm Air display ENTER air 5 Determine the weight of the sample in air Place sample on the balance 20 0 g 6 Store value for weight in air ENTER 7 Remove sample from the balance water 8 Determine we...

Page 41: ...decimal places minus1 Reduced by 1 place o Factory setting Features Set the basic unit and display accuracy in the Setup menu see Configuration Set the application weight units and display accuracies...

Page 42: ...on next page 3 Confirm the weight unit pounds ENTER pound o 4 Select the next application ENTER KEINE o weight unit in this example Troy ounces Repeatedly see table on next page MENU troy oz 5 Confirm...

Page 43: ...ains Grains 15 4323583500 GN 12 Peny Wt Pennyweights 0 64301493100 dwt 13 Milligr Milligrams 1000 00000000 mg 14 Pt P Lb Parts per pound 1 12876677120 o 15 Chn Tael Chinese taels 0 02645547175 tl 16 M...

Page 44: ...2 or ISO GLP compliant printout or record always on Setup Prnt Out GLP Alwayson menu code 1 6 7 3 Line format for printout include data ID codes 22 characters factory setting Setup Prnt Out Format 22C...

Page 45: ...Field for operator signature Blank line Dotted line ISO GLP compliant printout for external calibration adjustment Dotted line 17 Aug 2006 10 30 Date time beginning of measurement ACCULAB Balance manu...

Page 46: ...ed for printing out records of the results when you connect your balance to a Acculab data printer The record ing capability for printouts makes it easy for you to work in compliance with ISO GLP Data...

Page 47: ...ace port for printer Reset the menu factory settings output is blocked or Contact Acculab dealer err 02 Calibration parameter Calibrate only when zero not met e g is displayed balance not tared Press...

Page 48: ...analytical balances remove and clean the weighing pan as follows Reach beneath the drip breeze and lift it carefully together with the pan support to avoid damaging the weighing system Make sure that...

Page 49: ...al facility The packaging is made of environmentally friendly materials that can be used as sec ondary raw materials The equipment including accessories and batteries does not belong in your regular h...

Page 50: ...ibration weight AC power source power AC adapter 230 V or 115 25 V requirements voltage frequency 15 to 20 48 60 Hz Power consumption VA maximum 16 typical 8 STNG6 Approx hours of operation with the Y...

Page 51: ...pprox kg 4 4 4 7 Weighing pan size mm 90 d Weighing chamber height mm 230 Dimensions W D H mm 230 303 330 Model ATL 623 ATL 423 ATL 153 ATL 623 I ATL 423 I ATL 153 I Weighing capacity g 620 420 150 Re...

Page 52: ...F2 Operating temperature range 10 30 C 273 303 K 50 86 F Net weight approx kg 3 1 3 5 3 1 3 5 3 1 3 5 2 0 2 6 Weighing pan size mm 180 180 180 180 180 180 115 d Dimensions W D H mm 230 303 91 230 303...

Page 53: ...ing pan size mm 90 d Weighing chamber height mm 230 Dimensions W D H mm 230 303 330 Model ATL 623 V ATL 423 V ATL 153 V Type BD ED 200 BD ED 200 BD ED 200 Accuracy class 1 K K K Weighing capacity Max...

Page 54: ...weight approx kg 3 5 Weighing pan size mm 180 180 Dimensions W D H mm 230 303 91 Model ATL 822 V ATL 8201 V ATL 6201 V Type BD ED 200 BD ED 200 BD ED 200 Accuracy class 1 K K K Weighing capacity Max...

Page 55: ...2C connecting cable length 15 m 50 ft YSC01L15 Product Order No Density determination kit1 for ED224S ED124S YDK01LP Industrial AC adapter model ING2 protection rating IP65 in accordance with EN 60529...

Page 56: ...cations On request Sartorius will provide information on the minimum operating specifications in accordance with the Standards listed above for defined immunity to interference 73 23 EEC Electrical eq...

Page 57: ...57...

Page 58: ......

Page 59: ...rmission of Acculab The status of the information specifications and illustrations in this manual is indicated by the date given below Acculab reserves the right to make changes to the technology feat...

Reviews: