Example:
Determining animal weight with automatic start and 20 subweighing operations (measurements)
Parameter settings:
Applic. AnimalW.
(menu code 2. 7.)
Step
Key (or instruction)
Display/Data output
1. Place animal weighing bowl on the balance
22.6
g
2. Tare the balance
(TARE)
0.0
g
3. Change the number of subweighing operations:
(MENU)*
ref
30
4. Set number of measurements:
Repeatedly:
In increments of 1 (1, 2, 3, etc. to 100)
(MENU)* Press briefly
In increments of 10 (10, 20, etc. to 100)
or (MENU)* and hold
ref
20
5. Confirm number of measurements and start
(ENTER)
+
0.0
g
*
automatic animal weighing.
The number of measurements remains stored
in battery-backed memory until the setting is changed
6. Place first animal in bowl.
888
The balance delays the start of measure
20
ments until the difference between 2
19
measurements meets the criterion
....
1
$
If you selected the manual start mode, press
(ENTER)
7. Read off the result
+
410.1
g
G
*
The result is displayed with the
“*” symbol (= calculated value)
mDef
20
and remains displayed until the sample
x-Net +
410.1 g
(animal) is removed from the load plate (bowl)
8. Unload the balance
+
0.0
g
*
9. Weigh next animal (if desired)
Next weighing series begins automatically
* On some models, the keypad overlay shows the German word “MENUE“ for MENU.
34