Features
You can have device information, ID texts and date and
time printed before (GLP header) and after (GLP footer)
the values of a weighing series. These parameters include:
GLP header:
– Date
– Time at beginning of measurement
– Balance manufacturer
– Balance model
– Balance serial number
– Software version number
– Identification number of the current sampling operation
GLP footer:
– Date
– Time at end of measurement
– Field for operator signature
!
Operating the Balance with Printer:
(e.g. YDP03-OCE):
§
Select the following settings on the balance and on the
printer:
– Software handshake:
Setup interf. handshk. sftware
(menu code 1. 5. 4. 1)
Configuration
§
Setting menu codes for the printout (see “Configuration”):
– ISO/GLP-compliant printout or record only for
calibration/adjustment:
Setup Prnt.Out GLP Cal.-Adj.
(menu code 1. 6. 7. 2) or
ISO/GLP-compliant printout or record always on:
Setup Prnt.Out GLP Always on
(menu code 1. 6. 7. 3)
– Line format for printout: include data ID codes
(22 characters; factory setting):
Setup Prnt.Out Format 22 Char.
(menu code 1. 6. 6. 2)
– Formats for time:
setup prnt.0ut time 24 h
(menu code 1. 6. 8. 1) or
setup prnt.oUt time 12 h
(menu code 1. 6. 8. 2)
– Formats for date:
setup prnt.0ut date dd.mmm.yy
(menu code 1. 6. 9. 1) or
setup prnt.oUt date mmm.dd.yy
(menu code 1. 6. 9. 2)
!
No ISO/GLP-compliant record
is output if any of the following settings are configured:
Setup Prnt.Out Print Aut. W/o
or
Aut.With
(menu code 1. 6. 1. 3, 1. 6. 1. 4, ) or
Format 16 Char.
(menu code 1. 6. 6. 1)
Function Keys
Transfer header and first measured value: press (PRINT)
> The header is included with the first printout/data record.
To output header and reference data automatically when an
application program is active: press
Exit the application:
1) To send the GLP footer: press (CLEAR)
2) Quit application program: press (CLEAR) again
ISO/GLP-compliant Printout/Record
44