45
The ISO/GLP-compliant printout can contain the following lines:
--------------------
Dotted line
17-Aug-2006
10:15
Date/time (beginning of measurement)
ACCULAB
Balance manufacturer
Mod.
ATL-8201
Model
Ser. no.
10105355
Balance serial number
Ver. no.
00-36-01
Software version
ID
2690 923
ID
--------------------
Dotted line
L ID
Measurement series no.
nRef
10 pcs
Counting: reference sample quantity
wRef
21.14 g
Counting: reference weight
Qnt
+
567 pcs
Counting result
--------------------
Dotted line
17-Aug-2006
10:20
Date/time (end of measurement)
Name:
Field for operator signature
Blank line
--------------------
Dotted line
ISO/GLP-compliant printout for external calibration/adjustment:
--------------------
Dotted line
17-Aug-2006
10:30
Date/time (beginning of measurement)
ACCULAB
Balance manufacturer
Mod.
ATL-8201
Model
Ser. no.
10105352
Balance serial number
Ver. no.
00-36-01
Software version
ID
2690 923
ID
--------------------
Dotted line
Cal. Ext. Test
Calibration/adjustment mode
Set
+
5000.0 g
Calibration weight
Diff. +
0.2 g
Difference determined in calibration
Cal. Ext. Complete
Confirmation of completed calibration procedure
Diff.
0.0 g
Difference from target following adjustment
--------------------
Dotted line
17-Aug-2006
10:32
Date/time (end of measurement)
Name:
Field for operator signature
Blank line
--------------------
Dotted line