49
Safety Inspection
If there is any indication that safe
operation of the balance is no longer warranted:
§
Turn off the power and disconnect the equipment from AC
power immediately.
> Lock the equipment in a secure place to ensure that it can-
not be used for the time being.
Notify your Acculab dealer. Repair work must be performed
by trained service technicians.
We recommend having the power
supply inspected by a certified electrician at regular inter-
vals, according to the following checklist:
– Insulating resistance: > 7 megaohms measured with a
constant voltage of at least 500 volts at a 500 K-ohm load
– Leakage current: < 0.05 mA measured with a properly
calibrated multimeter
Information and Instructions on Disposal and Repairs
Packaging that is no longer required must be disposed of at
the local waste disposal facility. The packaging is made of
environmentally friendly materials that can be used as sec-
ondary raw materials.
The equipment, including accessories and
batteries, does not belong in your regular
household waste. The EU legislation
requires its Member States to collect elec-
trical and electronic equipment and dis-
posed of it separately from other unsorted municipal waste
with the aim of recycling it.
For disposal in Germany and in the other Member States of
the European Economic Area (EEA), please contact our ser-
vice technicians on location or your Acculab dealer.
In countries that are not members of the European Econom-
ic Area (EEA) or where no Acculab dealers are located,
please contact your local authorities or a commercial
disposal operator.
Prior to disposal and/or scrapping of the equipment, any
batteries should be removed and disposed of in local
collection boxes.
Equipment contaminated with hazardous materials (ABC
contamination) will not be taken back; neither for repair or
disposal.
Recycling