51
Specifications for Individual Models
Model
ATL-224, ATL-224-I ATL-124, ATL-124-I ATL-84, ATL-84-I
Weighing capacity
g
220
120
80
Readability
g
0.0001
0.0001
0.0001
Tare range (subtractive)
g
220
120
80
Repeatability
<
± g
0.0001 0.0001
0.0001
(std. deviation)
Linearity
<
± g
0.0002
0.0002
0.0002
Response time (average)
s
2.5
Sensitivity drift 10 to +30°C
<
±/K
2 · 10
–6
Adaptation
By selection of 1 of 4 optimized filter levels; display
to ambient conditions
update: 0.1 – 0.4 (depends on filter level selected)
External calibration weight
g
200 (E2)
100 (E2)
50 (E2)
Operating temperature range
+10 … +30 °C (273 ... 303 K, 50° ... 86 °F)
Net weight, approx.:
kg
4.4 | 4.7
Weighing pan size
mm
90
d
Weighing chamber height
mm
230
Dimensions (W
+
D
+
H)
mm
230
+
303
+
330
Model
ATL-623, ATL-423, ATL-153,
ATL-623-I
ATL-423-I
ATL-153-I
Weighing capacity
g
620
420
150
Readability
g
0.001
0.001
0.001
Tare range (subtractive)
g
620
420
150
Repeatability
<
± g
0.001
0.001
0.001
(std. deviation)
Linearity
<
± g
0.002
0.002
0.003
Response time (average)
s
1
1
1.3
Sensitivity drift within +10 to +30°C
<
±/K
2 · 10
–6
2 · 10
–6
5 · 10
–6
Adaptation
By selection of 1 of 4 optimized filter levels; display
to ambient conditions
update: 0.05 – 0.4 (depends on filter level selected)
External calibration weight
(of at least accuracy class...)
g
500 (E2)
200 (F1)
100 (F1)
Operating temperature range
+10 … +30 °C (273 ... 303 K, 50° ... 86 °F)
Net weight, approx:
kg
3.2 | 3.6
3.2 | 3.6
2.6 | 3.0
Weighing pan size
mm
115
d
Dimensions (W
+
D
+
H)
mm
230
+
303
+
136