52
Specifications for Individual Models
Model
ATL-6202,
ATL-4202,
ATL-2202,
ATL-822,
ATL-6202-I
ATL-4202-I
ATL-2202-I
ATL-822-I
Weighing capacity
g
6200
4200
2200
820
Readability
g
0.01
0.01
0.01
0.01
Tare range (subtractive)
g
6200
4200
2200
820
Repeatability
<
± g
0.01
0.01
0.01
0.01
(std. deviation)
Linearity
<
± g
0.02
0.02
0.02
0.03
Response time (average)
s
1.1
1.1
1.1
1.0
Sensitivity drift within +10 to +30°C
<
±/K
2 · 10
–6
2 · 10
–6
2 · 10
–6
5 · 10
–6
Adaptation
By selection of 1 of 4 optimized filter levels; display
to ambient conditions
update: 0.1–0.4 (depends on filter level selected)
(of at least accuracy class...)
g
5000 (E2)
2000 (E2)
2000 (F1)
500 (F2)
Operating temperature range
+10 … +30 °C (273 ... 303 K, 50° ... 86 °F)
Net weight, approx.:
kg
3.1 | 3.5
3.1 | 3.5
3.1 | 3.5
2.0 | 2.6
Weighing pan size
mm
180
+
180
180
+
180
180
+
180
115
d
Dimensions (W
+
D
+
H)
mm
230
+
303
+
91
230
+
303
+
87
Model
ATL-8201,
ATL-6201,
ATL-8201-I
ATL-6201-I
Weighing capacity
g
8200
6200
Readability
g
0.1
0.1
Tare range (subtractive)
g
8200
6200
Repeatability (std. deviation)
<
± g
0.1
0.1
Linearity
<
± g
0.3 | 0.1
0.3 | 0.1
Response time (average)
s
1
1
Sensitivity drift 10 to +30°C
<
±/K
10 · 10
–6
Adaptation
By selection of 1 of 4 optimized filter levels; display
to ambient conditions
update: 0.05 – 0.4 (depends on filter level selected)
External calibration weight
(of at least accuracy class...)
g
5000 (F2)
Operating temperature range
+10 … +30 °C (273 ... 303 K, 50° ... 86 °F)
Net weight, approx.:
kg
2.7 | 3.5
Weighing pan size
mm
180
+
180
Dimensions (W
+
D
+
H)
mm
230
+
303
+
91