19
Calibrate the Scale
Step
Key (or instruction)
Display
a. Switch on the scale
(ON/OFF)
b. Zero the scale
(ZERO/LEFT)
c. Select calibration/linearization mode
(TARE/RIGHT) > 2 Sec
d. Select calibration
(Fn/UP) or (G/N/DOWN) Repeatedly
e. Confirm/start calibration.
(PRINT/ENTER)
After the zero point is stored, the required
calibration weight is displayed
f. Place the required weight on the scale
g. If the weight is applied within the defined time (PRINT/ENTER)
limit and tolerance, the “OK” symbol is displayed.
Press “(PRINT/ENTER)” to confirm and save calibration.
(To exit calibration press (ZERO/LEFT) > 2 sec)
Calibration/Span Adjustment