21
Output Format with 22 Characters
The following data block format is output:
I
I
I
I
I
I
V
W W W W W W W
W W
U
U
U
CR LF
I : ID
W : Weight value
V : +/- sign
U : Unit
ID Codes
S t a t
Indicates a special weighing platform status, e.g., initialization, self-test, adjustment/calibration
N
Indicates net or gross value
N 1
Indicates net value with 1
st
tare memory assigned (not main tare)
N 2
Indicates net value with 2
nd
tare memory assigned (not main tare)
S / N
Indicates equipment serial number
M o d e l
Indicates equipment Model
Plus/Minus Sign
+
Plus sign
–
Minus sign
Space (i.e., no plus or minus sign; if weight value is 0 or if output does not include a weighed value)
Unit
No stable parameter.; no weighed value
g Grams
k g
Kilograms
l b
Pounds
o z
Ounces
t
Tons
Special Codes
The data block may contain special information.
Special status-dependent codes:
1
2
3
4
5
6
7
8
9
10 11 12 13 14 15 16 17 18 19 20 21 22
S1 S2
CR LF
The following status codes are output for “S1” and “S2”:
: Taring
C : Internal calibration
- - : All numerals shown in stable readout
H : Overload
L : Underload
SBI Interface Protocol Description