background image

ENGLISH

4

Digital Calibration:

Power Up Test:

Overload:

Auto Off:

RS-232:

Allows for convenient user calibration from the keypad with three selectable weight options.

When the unit is initially turned on all segments of the display will appear. The segments will be displayed for 
approx. 3 seconds. After self-test scale will return to zero.

When a load or force is applied to the scale platform that exceeds the unit’s capacity “H” will appear in the 
display indicating an overload has occurred. 

During battery operation (in order to conserve battery consumption) the scale will automatically turn off after 
approximately 4 minutes of inactivity / default setting. (Please see “Menu” options for additional settings).

Standard feature included   with scale. (Please see Interface Description)

Features:

Care and Maintenance:

Avoid lengthy exposure to extreme heat or cold. Please allow the scale to acclimate to room temperature 
for at least 15 to 20 minutes if exposed to extreme temperatures. (Please see Product Specifications)
Keep product clean. Remove pan from base and clean accordingly. Do not “wash down” scale or indicator 
with water or cleaning liquids.
Please use supplied packaging when transporting scale.
Remember to remove battery when not in use.

1

2

3

5

4

Totaling:

To select the application “Totaling” press and hold “TARE/ZERO” key for more than 3 seconds to access 
application menu.
With the “F” key you can scroll through the menu options (UNIT, COUNT, & TOTAL).
Select “TOTAL” by pressing the “CAL/PRINT” Key.
Display will automatically return to weighing mode with “TTL” displayed (signifying to the end user that 
they are in application mode).
Place1st object on scale and press “F” (value is stored) / remove object and display will return to zero.
Place 2nd object on scale and press “F” (value is added to 1st) and the “TTL” symbol will flash (indicating 
total weight stored) / remove object and display will return to zero (“TTL” stops flashing). 
Place 3rd object on scale and press “F” (value will be added and “TTL” will flash) / remove object and 
display will return to zero (”TTL” stops flashing) process can be repeated up to 99 times.
Holding down the “F” key for 3 seconds will clear all memory.
To exit “Totaling”, press and hold “TARE/ZERO” Key for more than 3 seconds to access application menu 
and select “Unit” (default setting).

6

8

7

9

To exit “Counting” press and hold “TARE/ZERO” key for more than 3 seconds to access application 
menu and select “UNIT” (default setting).

9

Summary of Contents for SVI-100E

Page 1: ...PROFESSIONAL PRECISION INDUSTRIAL SCALE SVI SERIES SVI 50C SVI 100E SVI 200F SVI 10A SVI 20B...

Page 2: ...ion Programs Features Care Maintenance Menu Settings Scale Operating Manu Interface Description Product Specifications AC Adapter Battery Operation On Off Cal Print F Key Zero Tare Light Weight Unit C...

Page 3: ...recommended to remove battery when scale is not in use for extended periods Also replace low or dead batteries Note Back lit display is off unless activated with Light key ACAdapter Battery Operation...

Page 4: ...cription Calibration Weight Calibration with the pan cleared and zero stability indicated in display press and hold the CAL PRINT key for more than 3 seconds to enter the calibration mode Depending up...

Page 5: ...t options g lbs ozt kg depending upon model Select desired weigh unit by pressing CAL PRINT selected option will be displayed and balance will return to weighing mode Pressing the F key will enable yo...

Page 6: ...th water or cleaning liquids Please use supplied packaging when transporting scale Remember to remove battery when not in use 1 2 3 5 4 Totaling To select the application Totaling press and hold TARE...

Page 7: ...ore selected codes and exit menu press and hold ZERO TARE key for 3 seconds or until display returns to weigh mode 6 8 7 9 Unit Toggling Auto Zero Calibration Setting Baud Rate Parity 141 Unlocked 0 1...

Page 8: ...Totaling Mode Calibration Mode ASCII p t u f a b c D HEX 70 0D 0A 74 0D 0A 75 0D 0A 66 0D 0A 61 0D 0A 62 0D 0A 63 0D 0A 64 0D 0A CR LF CR LF CR LF CR LF CR LF CR LF CR LF CR LF Output Format Position...

Page 9: ...ozt 100 kg 220 lbs 3215 ozt 200 kg 440 lbs 6430 ozt 1 g 0 001 kg 0 002 lbs 0 05 ozt 2 g 0 002 kg 0 005 lbs 0 1 ozt 0 005 kg 0 01 lbs 0 2 ozt 0 02 kg 0 05 lbs 0 5 ozt 0 05 kg 0 1 lbs 1 ozt grams kilog...

Page 10: ...Zone 100102 Bejing PR China TEL 86 10 6439 2233 If you have any questions please contact your local Acculab supplier visit our website www acculab com or contact Si tiene alguna duda p ngase en contac...

Reviews: