When the probe is inserted after the SpO

2

main unit is turned on, the state

indication is “Testing...”.

Summary of Contents for WS Series

Page 1: ...Copy Right Hunan Accurate Bio Medical Technology Co Ltd Page 1 Total 29 User Manual WS Series Pulse Oximeter Please read the manual carefully before use...

Page 2: ...Hunan Accurate Bio Medical Technology Co Ltd Without the written permission of Accurate no individual or organization shall reproduce amend or translate any part of this manual Version of Manual V1 1...

Page 3: ...t having the most complete configurations Therefore some content hereof may not apply to the product you have purchased If you have any question please feel free to contact us Please keep this manual...

Page 4: ...oduct Function List 6 3 Preparation before Use 8 3 1 Unpacking Inspection 8 3 2 Environmental Requirements 8 3 3 Start up 8 3 4 Shutdown 8 4 Basic Operations 10 4 1 User Setup 10 4 2 System Setup 11 4...

Page 5: ...Co Ltd Page 5 Total 29 7 1 Check 17 7 2 Cleaning and Disinfection 17 7 3 Scrapping 18 8 Troubleshooting 19 A Product Specifications 20 Technical Description 21 B Network Security 23 C EMC 24 D Default...

Page 6: ...nvironment with anesthetic agent or other inflammables or explosives Do not open the housing of the device In case of any problem please contact your dealer or the manufacturer The patient s safety sh...

Page 7: ...ould result The temperature of the enclosure such as the button and LCD display can reach 44 4 degrees The internal temperature of the shell near the battery can reach 47 1 degrees So don t touch butt...

Page 8: ...s been developed in accordance with the requirements of IEC 60601 1 4 to minimize the probability of risks caused by program error Do not attempt to open the housing for repair If the product is damag...

Page 9: ...separate collection after its useful life Non ionizing radiation Use by date Temperature limit No alarm system IP22 Degree of protection against liquid ingress Caution consult accompanying documents A...

Page 10: ...se under hypoperfusion and exercise conditions 2 1 2 Intended Use The Pulse Oximeter is intended to measure functional arterial oxygen saturation SpO2 and pulse rate of patients in Home Health Care an...

Page 11: ...nfirm key 3 Up key This key has different functions in different situations Press this key to move the cursor upward increase the value of a menu item etc 4 Down key This key has different functions i...

Page 12: ...opy Right Hunan Accurate Bio Medical Technology Co Ltd Page 7 Total 29 Rechargeable Bluetooth Storage represents that the device has this function represents that the device does not have this functio...

Page 13: ...isms during storage transport and use Please confirm the package is complete prior to use and do not use the device if any damage is found 3 2 Environmental Requirements The operating environment for...

Page 14: ...shut down the oximeter according to the following steps 1 Confirm that the measurement is to be ended 2 Disconnect the SpO2 probe from the oximeter 3 Place the main unit of oximeter still for a while...

Page 15: ...Sex 2 Press the Up Down key for selection 3 Press the Confirm key to confirm the modification and return to other menu for selection SpO2 Reminder ON OFF 1 Press the Confirm key to confirm the select...

Page 16: ...e Up Down key to select the setting item and press the Confirm key to confirm the selection 3 After selecting the item press the Up Down key to change the value and press the Confirm key to confirm th...

Page 17: ...t on the display screen When the battery level is displayed as please charge the battery timely 4 6 Transmission via Bluetooth Turn on Bluetooth on your smart phone launch the specific application to...

Page 18: ...Copy Right Hunan Accurate Bio Medical Technology Co Ltd Page 13 Total 29 When the probe is inserted after the SpO2 main unit is turned on the state indication is Testing...

Page 19: ...tion state 3 Battery level 4 SpO2 The percentage of oxyhemoglobin in total hemoglobin 5 PR The detected number of pulses per minute 6 Bar chart The amplitude of bar chart represents the level of pulse...

Page 20: ...equipment should not be entangled with the cable of SpO2 probe Do not place the SpO2 probe on a limb with any arterial duct or intravenous line Do not place the SpO2 probe and the BP cuff on the same...

Page 21: ...r end of the charging cable to the charger 3 Disconnect the adapter after full charging NOTE The service life of lithium battery depends on the time and frequency of use If the lithium battery is main...

Page 22: ...device please power it off and disconnect the charging cable and SpO2 probe CAUTION If any liquid is poured onto the device or its accessories by accident please contact the service personnel or our C...

Page 23: ...t the device 7 3 Scrapping To avoid contaminating the environment or other equipment or infecting other people please disinfect and purify the device and its accessories according to applicable nation...

Page 24: ...Short press the Power key if the oximeter cannot be turned on it indicates the oximeter is damaged Please contact the local customer service center 3 If the oximeter can be turned on it indicates the...

Page 25: ...ut providing explosion protection Rating of protection against liquid ingress IP22 Rating of movement Wrist type Working mode Continuous Physical specification Width Height Thickness 47 55 17 mm Max w...

Page 26: ...Description The table below shows the statistical conclusion of the study on invasive controlled desaturation according to Annex EE Guideline for evaluating and documenting SpO2 accuracy in human obje...

Page 27: ...Copy Right Hunan Accurate Bio Medical Technology Co Ltd Page 22 Total 29 Bland Altman Plot for SaO2 SpO2 Standard deviation of SaO 2 SpO 2...

Page 28: ...The mobile phone management software for the pulse oximeter The operating environment is a mobile phone installed with Android 8 0 OS or higher compatible version Minimum hardware configuration requir...

Page 29: ...elow Other devices may affect this pulse oximeter even though they meet the requirements of CISPR When the inputted signal is below the minimum amplitude provided in technical specifications erroneous...

Page 30: ...ines 1 kV signal input output 100 kHz repetition frequency 1 kV signal input output 100 kHz repetition frequency N A Surge IEC 61000 4 5 0 5 kV 1 kV differential mode 0 5 kV 1 kV 2 kV common mode N A...

Page 31: ...apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for rad...

Page 32: ...es objects and people Recommended separation distances between RF wireless communications equipment The device is intended for use in an electromagnetic environment in which radiated RF disturbances a...

Page 33: ...e equation applicable to the frequency of the transmitter Recommended separation distance where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufact...

Page 34: ...SpO2 low reminder 90 Upper limit for PR reminder 120 Lower limit for PR reminder 40 E Product and Accessories No Item Quantity 1 Main unit 1 2 SpO2 probe 1 3 User Manual 1 4 Charging cable 1 The main...

Page 35: ...Copy Right Hunan Accurate Bio Medical Technology Co Ltd Page 29 Total 35...