2. Product Overview
2
7
6.5 Maximum Power Limit
The myVolt Touch Power Supply has a maximum power output
of 100 Watts.
Per Ohms Law:
Power (P) = Current (I) x Voltage (V)
If the maximum power is reached during an electrophoresis run,
the non-constant parameter will be limited to prevent exceeding
the maximum power output
.
CAUTION:
It is not recommended to run the power supply at or
near the maximum wattage, as this will cause the power supply
and gel tank to get hot. Most standard electrophoresis buffer
concentrations will not have such low resistance that the current
and voltage level will come close to the maximum power limit.
Touch Screen
Control and Display
4 Pairs of
Electrical Outputs
Folding
Front Feet
Recesses for
Stacking 2 Units
AC Power Inlet
and Fuse
6.4 Automatic Cross Over
During an electrophoresis run, the power supply will attempt to
maintain the constant voltage or amperage setting, depending on
which parameter is set to “CONSTANT”.
If, during a run, the non-constant parameter setting (limit) is met
due to the resistance of the load, the power supply will
automatically “cross over” so the non-constant parameter will
now be the constant and limiting parameter. A red outline will
appear around the new “constant” value.
FRONT
REAR
Rating, Serial #
Label
In CONSTANT CURRENT mode, the power supply will output
the set mAmp current parameter, and the voltage will fluctuate
relative to the resistance of the load (buffer concentration in the
attached electrophoresis tanks) up to a maximum of the voltage
setting that has been chosen.
Cooling
Fan