background image

3

3.  myVolt Touch Specifications

myVolt Touch, Model E2301, 300V Power Supply

Output Voltage

10 to 300 VDC, in 1 V increments

Voltage Accuracy

+/- 2% of displayed value or 3V 

(whichever is greater)

Output Current

4 to 400mA, in 1 mA increments

Current Accuracy

+/- 2% of displayed value or 5mA 

(whichever is greater)

Maximum Power

100 Watts 

Output Terminals

4 Pair, wired in parallel

Timer

1min to 9 hours, 59 minutes

Control Panel

Capacitive touch screen

Display

Color TFT, LCD 

Fuse

250V 2.0A, 5x20mm,Type T

Operating Temp

– 40

°

C

Humidity

– 95%

Dimensions

8.5x6.3x4.0in/21.5x15.5x9.75cm(W

×

D

×

H)

Net weight

1.5 lbs / 0.68 kg

Input Electrical

100 to 240 VAC, 50/60Hz

6

6.3  Constant Voltage / Constant Amperage

The myVolt Touch can be set to run with constant voltage or 
constant amperage output, with the non-constant parameter set as a 
limit.   

Press the box “CONSTANT …” to switch between “CONSTANT 
CURRENT” and “CONSTANT VOLTAGE” running modes.

In constant voltage mode, the power supply will output the set 
voltage, and the mAmp current value will fluctuate (“float”) depending 
on the resistance of the load, up to a maximum of the mAmp current 
setting that has been chosen.  The buffer concentration and size of 
the electrophoresis gel box will dictate the resistance.

6.2  Setting Parameters

The myVolt Touch Power Supply has 3 main parameters that can 
be adjusted: Run Time (TIME), Voltage (VOLTAGE), and Current 
(mAMP).

(See section 7 for an overview of Voltage, Current, Resistance, and 
Power)

Press the parameter box to adjust the value.  The selected 
parameter box will be highlighted.  Press the UP/DOWN arrows to 
change the value.

Adjust parameters (TIME, VOLTAGE, mAMP) as required for the 

electrophoresis application.

Summary of Contents for E2301 Series

Page 1: ...300V Power Supply Operating Manual This product is for research only Service and Contact Series E2301 A division of Version 3 18 Software rev 1 8 ...

Page 2: ...s that could damage plastic Use caution when cleaning near the housing vents Do not allow liquid to enter the power supply 9 Troubleshooting Problem Causes Measures No display Power not connected Fuse blown Check cord connection and power supply Unplug and re attach power cord Check and replace fuse at power inlet Voltage or Amperage setting not reached Maximum wattage exceeded Resistance of the l...

Page 3: ...nks that incorporate sufficient safety features to prevent electric shock Some types of shielded plugs may not fit into the output jacks and in this case an adapter 1 7 Overview of Electrical Principles Definitions Voltage The amount of potential energy between two points in a circuit Current The time rate of flow of an electric charge measured in amperes Resistance A material s opposition to the ...

Page 4: ...trical Outputs Folding Front Feet Recesses for Stacking 2 Units AC Power Inlet and Fuse 6 4 Automatic Cross Over During an electrophoresis run the power supply will attempt to maintain the constant voltage or amperage setting depending on which parameter is set to CONSTANT If during a run the non constant parameter setting limit is met due to the resistance of the load the power supply will automa...

Page 5: ...t voltage or constant amperage output with the non constant parameter set as a limit Press the box CONSTANT to switch between CONSTANT CURRENT and CONSTANT VOLTAGE running modes In constant voltage mode the power supply will output the set voltage and the mAmp current value will fluctuate float depending on the resistance of the load up to a maximum of the mAmp current setting that has been chosen...

Page 6: ... water direct sunlight strong light corrosive gas high magnetic fields heaters fires and other heat sources 4 5 6 Operating guide 6 1 Touch Screen Control Up Down Value Adjust Buttons Adjustable Parameters Operation Buttons Status 4 Installation Place the power supply on a smooth level surface in an area where it will not contact liquids or chemicals The myVolt Power Supply features Dual Orientati...

Reviews: