7.5 Setting the weighing parameters
For proper use of the balance, it is important to make the correct weighing parameter settings
according to the operating environment.
This section describes the filter parameters, stability, autozero and weighing units.
In the settings menu, select the "
Setup
" function and then the "
Weighing parameters
"
function.
In the "
Weighing parameters
" screen, the set
values (in the box to the right of the parameter)
are immediately verifiable.
By touching the parameter key you want to
change, you can change the value.
▪
Filter:
the filter function allows to speed or
slow down the balance response according to
weighing requirements and environmental
conditions. By selecting Level1, the response
will be immediate, but the balance will remain
more sensitive to environmental disturbances
such as ventilation and vibration. By
increasing the level, the response will be
slower and the indication more stable.
➢
Level 1:
dosage conditions
➢
Level 2:
stable conditions
➢
Level 3:
unstable conditions
▪
Stability:
this function allows you to adjust the
balance to the working environment. When
using
the
device
in
a
vibration-free
environment, select level 0. The default level
is 2. Use level 3 for very disturbed
environments.
➢
Level 0:
For stable environments.
➢
Level 1:
For poorly stable
environments.
➢
Level 2:
For unstable environments.
➢
Level 3:
For heavily unstable
environments.