51
8.4
“
Weight %
”
function
This function allows you to determine the weight percentage of a sample based on a
reference weight.
The value of the reference weight can be obtained by a sample weight or manually entering
the weight value.
To access the function from the "Functions" menu, tap the "
Weight %
" icon.
Select the mode:
Automatic reference:
Selecting this mode upon startup will
require you to load the reference sample
weight (100%) on the weighing plate.
Manual reference:
Selecting this mode requires you entering
the reference weight value (100%). Then
tap the "
Reference Weight
" key and enter
the weight value in the input screen and
select the desired unit of measurement.
Start:
after selecting the desired mode and
entering the required data, tap the "Start"
button to run the "Weight%" function.
Database:
by tapping the "Database" button, you can
create a custom database to perform
weighing
in
percent
with
different
methodologies.
Using the database in addition to the
simple percentage weighing, you can
perform statistical weighing checks or set
up to two levels of weight check thresholds.
Tap the "
New
" key and enter the name of
the new record to be stored.
▪
Description:
enter a description of the
item.
▪
Mode:
select the weight reference
mode.
Summary of Contents for ES59135
Page 7: ...6 ...