background image

 

69 

 

8.9   

Maximum load

 function 

 
This function allows to measure the maximum breakage load of a solid. 
 
To access the function from the "Functions" menu, tap the "

Maximum load

" icon. 

 
 

 

 
 
 

 

Selecting the function immediately displays 
the measurement screen. 

  Prepare  the  sample  to  be  tested  on 

the weighing plate. 

  Perform  tare  or  zero  reset  if 

necessary. 

  Tap the "

Ok

" key to start the test. 

 
 
 
 
 
Once  the  tested  sample  breakage  load  is 
reached,  the  maximum  weight  will  be 
displayed. 
 
 
Weight detected on the plate 
 
 
 
 
 

 
Tap the "

Esc

" key to exit the function, the "

Restart

" key to make a new measurement and the 

"Print"

 key to print the measurement result. 

Summary of Contents for ES59135

Page 1: ...Operation Manual Analytical Balance Series Tx ES59135 Models W3002A 120 W3102A 220 Touch Screen Revision 1 18 ...

Page 2: ...TING THE WEIGHING PARAMETERS 29 7 6 SELECTING THE UNIT OF MEASUREMENT 30 7 7 SELECTING THE CUSTOM MEASUREMENT UNITS 31 7 8 CALIBRATING AND SETTING THE CALIBRATION MODE 32 7 9 SETTING THE PERIPHERALS 35 7 10 INFO ABOUT THE SOFTWARE 39 7 11 SERVICE 39 7 11 1 System data backup and restore 39 7 11 2 Factory reset 41 7 11 3 Software update 42 7 11 4 Balance technical service 42 8 FUNCTIONS MENU 43 8 1...

Page 3: ...64 8 7 2 Screen of the database textile function with statistics enabled 65 8 8 ANIMAL WEIGHTING FUNCTION 67 8 8 1 Screen of the Animal weighting function 68 8 9 MAXIMUM LOAD FUNCTION 69 8 10 RECIPE FUNCTION 70 8 10 1 Screen of the Simple recipe function 73 8 10 2 Screen of the Database recipe function 74 8 11 SETTING THE GLP DATA 76 9 TECHNICAL FEATURES 78 10 WARRANTY 79 11 DISPOSAL 79 ...

Page 4: ...erly used the warranty can be void Do not use the device in areas at risk of explosion Before starting the device for the first time check if the power supply unit or the power cord is damaged and confirm that the voltage of the wall outlet to be used corresponds to the mains voltage required 2 Storage conditions Storage temperature 5 C 40 C Storage humidity 45 75 Keep the original balance packagi...

Page 5: ...ing Power supply unit Power supply box Remove the accessory box and then carefully lift the balance with the protective foam parts in place IMPORTANT Keep the original balance packaging in case the instrument needs to be shipped for servicing disconnect all cables remove weigh pan and pan support place rubber band around draft shield or tape the glass doors closed and separate and package accessor...

Page 6: ...rized Service Center to prevent damage Below are the procedures Before packing packing remove the weight pan pan support draft ring and any other loose parts and place them in the accessory box Following the below steps to pack again the balance in its box User Manual 1 2 3 4 ...

Page 7: ...6 ...

Page 8: ...7 3 2 Overview and accessories installation 1 Pan Support 2 Weigh Pan 3 Power supply unit 2 1 3 ...

Page 9: ...aging check for visible damage to the device Do not install the balance in any room or location where there are any air currents strong heat shifts or vibrations Do not use the balance in explosive atmospheres The ambient humidity of the environment should be between 45 and 75 ...

Page 10: ...es allow the balance to equilibrate for 12 hours For calibration procedures please refer to the Device calibration chapter Perform the device calibration whenever it is moved to another place Never drop objects onto the the balance weigh pan and never place objects on the pan that weight more than the maximum capacity for the balance Any service must be performed by trained personnel and the spare...

Page 11: ...rations When switching off wait for the green LED to turn off before removing the power supply socket if necessary 4 Serial interface connections The device is equipped with an RS 232C interface for connection to a serial printer To ensure communication between the device and a printer the following conditions must be met Connect the printer using a suitable cable matching the position of the prin...

Page 12: ...eatures of the device 1 Weighing value display g Measuring unit indication by tapping the symbol you can access the measurement unit menu and select the desired unit Stability indicator T Man Indication of weight displayed Net net value Ldr gross value T Tare value indicator Analog indicator bar of magnitude being measured Technical Specifications for the instrument 2 3 1 ...

Page 13: ... on or off menu for the device standby mode Tap this area to access the user menu 3 Function key menu bar Touching the keys displayed in this area you will run the corresponding function Below are all the keys available in the simple weighing screen The function key display bar varies according to the selected applications Key bars available in simple weighing screens Print Data printing activatio...

Page 14: ...device Mod 1 Mode 2 Mod 3 Alphanumeric keypad Selection key for entering uppercase or lower case letters Delete key of the last entered character Delete key for all entered characters Back and forward cursor movement key on the inserted string Key to confirm and store the entry Keyboard character selection key Tap this key to rotate the various characters available for entry Mod 1 Mod 2 Mod 3 Inpu...

Page 15: ... Numeric keypad Delete key of the last entered character Key to confirm and store the entry Delete key for all entered characters Input operation cancellation key Key for selecting the unit of measurement ...

Page 16: ... the display will be off and it will remain ON only the lower right command to turn on the system again Touch the symbol to return to the basic weighing screen Logout Select the logout mode to enable the user used Touch the key Start to enable again the user and if the user is password protected you will be required to enter it Touching the key you can select another user and start the device agai...

Page 17: ...en At first power on the device starts by activating the default user with the factory settings In the weighing screen the area displaying the weight shows the value of the weight on the weighting plate Use the 0 key to set a new zero point and reset all tare values If you work with weighing boxes first zero the balance Place the container and press the T key Or use the T Man key to enter a known ...

Page 18: ...ility symbol to light up before detecting the value Press Lrd and Net to display the gross and net weights and vice versa To reset the balance remove all items from the plate and press the 0 key If you entered a manual tare press T Man and C key to clear the tare value ...

Page 19: ... Calibration menu Setup Users setup Preferences Info Service To access the settings menu tap the button in the top bar of the display The following screens will be displayed with all the functions of the setup menu Indicator of the number of pages available To scroll through the next and back pages tap on the arrows Return to the previous screen Exit from the menu Function indicator not available ...

Page 20: ...unction Tap Language in the menu list You can select the device s use language Then select the desired language Automatically all screens will be translated into the selected language N B The selected language will only be assigned to the user with whom the selection was made ...

Page 21: ...it will be changed for all users Tap Date and time in the menu list Tap Date in the menu list to change the date Tap Time in the menu list to change the time Tap 24h 12h in the menu list to set the time format A numeric keypad will be displayed for entering the date and time in dd mm yyyy mode for the date and hh mm ss for the time Tap OK to confirm the new settings C to delete all entered data or...

Page 22: ...rator Standard Creating an administrator user allows the user to access all the device setting features and changing or deleting other users Creating a standard user allows the user to access only some of the device setting features The prohibition display that appears when you touch a feature that is not available for the standard user is as follows You must log on with an administrator user to c...

Page 23: ...ters allowed is 30 The insertion method distinguishes between uppercase and lowercase letters so a user entered as TEST will be different from a user entered as a test After searching for a new user go back to the main screen Tap the user selection area and select the user you have just created From now on all changes to the functions databases or analytics results will be associated with the user...

Page 24: ...ect Administrator to set the user with administrator rights Settings Selecting this function will prompt you to enter a user s password with administrator rights Then select the Enable function to temporarily unlock the standard user or select the Disable function to lock the user again After you unlock the user in the base screen on the left side of the username an open padlock will appear Unlock...

Page 25: ...ate a new user with the same configuration as an existing user Any databases will not be copied and no password protection will be enabled To copy the databases of another user you need to export and import them to the desired user Select the user you want to duplicate Enter the name to assign to the new user and confirm with the Enter key Delete users By selecting the Delete Users function you ca...

Page 26: ...rd You will then be asked to re enter the password for confirmation Enter it and tap the OK key From this moment on every time you want to access the user you will be required to enter your password N B Write down your password and store it in a safe place Warning In case of loss you will only be able to recover the password for the Standard user The recovery procedure is described in chapter 8 3 ...

Page 27: ...d After selecting User Password you will be asked to enter the Old Password Enter an administrator password Then enter the New password to assign to the user Confirm the password you have just entered From now on the Standard user will have a new login password N B To remove the password upon the request for entering the new password and confirmation of the new password tap the OK key without typi...

Page 28: ...rameters that can be set in the Preferences menu are as follows Language see chapter 7 1 Appearance Brightness Beeper Functions Appearance You can choose between 6 different color combinations and select the desired one by tapping the corresponding key Brightness There are 4 levels of brightness on the display Select the desired one by tapping the corresponding key Beeper The beeper function enabl...

Page 29: ...o the sequence assigned to function keys F1 to F10 To assign a function press the key you want to assign and select one from the list of available functions To remove a function press the key corresponding to the function you want to remove and select None in the list From the base weighing screen tap to access the functions menu To switch to the next or previous screen tap the keys Next Back ...

Page 30: ...slow down the balance response according to weighing requirements and environmental conditions By selecting Level1 the response will be immediate but the balance will remain more sensitive to environmental disturbances such as ventilation and vibration By increasing the level the response will be slower and the indication more stable Level 1 dosage conditions Level 2 stable conditions Level 3 unst...

Page 31: ...ment Below is shown how to select the desired weighing unit In the settings menu select the Setup function and then the Units selection function On the Units Selection screen all available weighing units are immediately displayed Tap the desired unit to select it and enable it The last three units of measure K1 K2 K3 are customizable units For reference refer to the chapter Selecting the custom me...

Page 32: ...tup screen you can select one of the three units K1 K2 K3 available to personalize it After the selection enter the following data Name enter the unit identifier max 15 characters Symbol enter the symbol you want to assign to the unit of measurement max 3 characters The weight units g kg lb and so on are not allowed Multiplier factor enter the multiplicative factor with which the weighing result h...

Page 33: ... function in the settings menu In the calibration menu you can calibrate display and print the data for the last calibration and set the calibration mode Before performing the calibration procedure check that the plate is empty Calibration By tapping the calibration key the calibration procedure corresponding to the set calibration mode will be activated Load the required calibration weight or tap...

Page 34: ... Calibration mode the balance can be set to calibrate with two different modes External default By selecting the default calibration mode when the calibration command is given the required weight will be the factory default setting the value of which will vary depending on the balance model External custom By selecting the external custom calibration mode you can set a custom calibration weight va...

Page 35: ...alibration mode the balance automatically calibrates when the ambient conditions change or within a set time of 60 to 999 minutes Also with this mode you can calibrate each time you want by tapping the Calibration key To change the auto calibration range tap the Calibration Range key and set the desired range between 60 and 999 minutes The internal and automatic calibration mode is only available ...

Page 36: ...box on the right of the parameter Tapping the parameter key to be set will activate its screen with the list of the various selectable options Printer this function allows you to select the device connected to the serial output Dpp250 command print by touching the Print button for the Dpp250 printer Dpp350 command print by touching the Print button for the Dpp250 printer Tlp50 command print by tou...

Page 37: ...ey enabled mode before executing the print command verify that the save USB stick is inserted in the USB port on the left side of the device While saving the file on the screen you can scroll through the information that will be saved in the file At the end of the save the name of the file assigned automatically will appear on the screen The USB stick will then contain a compressed file with a tex...

Page 38: ...er at the beginning of the receipt The function can be On Bill heading enabled Off Bill heading disabled Print heading Setup Activation of the screen for inserting the desired text for the receipt header In the print header screen tap the rectangles that indicate the line number to automatically enable the alphanumeric keypad and enter the desired text with a maximum number of 40 characters per li...

Page 39: ...ng after user have entered it On g l p data print enabled Off g l p data print disabled For information entry and G l p database management see chapter 9 9 Time and date this function enables you to enable or disable the time and date printing On Calibration data print enabled Off Calibration data print disabled Time and date printing cannot be deactivated with the G l p function enabled Balance c...

Page 40: ...e and Backups of the device data In the settings menu select the Setup function then the Service one The functions available in the Service menu are as follows System data backup and restore Factory reset Software update Balance technical service 7 11 1 System data backup and restore You can back up the system completely or restore data and configuration previously saved on a USB stick This sectio...

Page 41: ...eport that the export operation was successful A backup folder with the assigned name was then created in the USD stick Opening the folder you will find a set of files and subfolders that contain all the data N B Do not modify or rename any files to avoid damaging the content The folder will be automatically restored to the device by using the System Restore command This folder will be automatical...

Page 42: ...he backup you want to restore A screen with the message Completed will report that the system recovery operation was successful 7 11 2 Factory reset You can perform a total reset of the device to bring it back to the initial conditions Tap the Factory Reset button confirm by touching the Yes button or cancel the operation by tapping the No button N B Be careful as all the data of all users includi...

Page 43: ... upgrade program supplied by the manufacturer and confirm with the Ok key If the selected file is correct the upgrade operation will be enabled and a displayed counter will indicate the progress status N B Once the software upgrade operation has begun do not power off the device until the operation completes During the upgrade the device will turn off and restart automatically do not press any key...

Page 44: ...d and sequenced according to its usage needs See section on preferences in the functions section So each time you change the user its function settings will vary Tap the button to access the functions menu Use the keys to scroll through the displayed back and next pages Tap the icon corresponding to the function you want to access the relevant settings To exit the functions menu tap the button or ...

Page 45: ...ing this key you can call up a program by entering the corresponding name Selection list by tapping this button you can call up a program by selecting it from a list New insert new record refer to the paragraph corresponding to the desired function Edit by tapping this button you can access the editable parameters of the function Delete tap this button to delete one or more previously stored recor...

Page 46: ...umber of pieces Enter the number of samples you want to use for the count by tapping the Number of samples key Unit mean weight Count by entering the weight value of the single piece to be counted Enter the weight value by tapping the button Unit mean weight Start After selecting the count mode tap the Start button to perform the count of the pieces Carry out the container tare if necessary in the...

Page 47: ... a description of the item Mode select the count mode Samples number Unit mean weight enter the number of samples or the average unit weight of the pieces Options you can select the statistical control by activating the Statistics function or by controlling the pieces by activating the Control of value function Statistics To enable this function you need to include Nominal value enter the nominal ...

Page 48: ...e number of pieces is within the set range Save tap the Save button to store the recorded record 8 2 1 Pieces count screen with Statistics function The following describes the pieces counting functions with statistical calculation Tap the Ok key to capture a new weight and tap the End key to end the capture and perform the final statistical calculation Parameters used for counting Statistics Weigh...

Page 49: ...ing the i key you can see the parameters used for counting the pieces The pieces number control bar is yellow when the number of pieces loaded is lower than the minimum threshold It is green when the pieces are within the set range with the beep indication when activated and red when the number of pieces exceeds the upper threshold set ...

Page 50: ... measurement to be used for testing by tapping the Unit key To deactivate the limit enter the value 0 or tap the C key and then tap the OK key Beeper enable or disable the beeper In case of enabling when the weight value is within the set range a beep is release Start After entering the required data tap the Start button to activate the weight check function Database By tapping the database button...

Page 51: ... i key you can see the parameters used for counting the pieces The Check weight bar is yellow when the loaded weight is lower than the minimum threshold and green when the weight is within the set range with the beep indication when activated and red when the weight is greater than the upper threshold set ...

Page 52: ...reference weight value 100 Then tap the Reference Weight key and enter the weight value in the input screen and select the desired unit of measurement Start after selecting the desired mode and entering the required data tap the Start button to run the Weight function Database by tapping the Database button you can create a custom database to perform weighing in percent with different methodologie...

Page 53: ...ue Enter the nominal weight value to be checked T2 T2 T1 T1 enter the error tolerance threshold values Optional entry Save tap the Save button to store the recorded record Check weight To enable this function you need to enter Lower limit enter the minimum acceptable weight value Upper limit enter the maximum acceptable weight value Sound you can activate or deactivate the beeper function when the...

Page 54: ...functions of the weighing with statistical calculation Tap the Ok key to capture a new weight and tap the END key to end the capture and perform the final statistical calculation Parameters used for the weight Weighting statistics Weighting performed Last weighing delete key ...

Page 55: ...heck By tapping the i key you can see the parameters used for weight check The weight check bar is yellow when the loaded weight is lower than the minimum threshold green when the weight is within the set range with the beep indication when activated and red when the weight is greater than the upper threshold set ...

Page 56: ...selecting this type of measure you will be required to enter the solid density value g cm3 If using distilled water or ethanol select the relevant item below and enter only the liquid temperature value The density of the selected liquid at the temperature entered will automatically be entered The reference tables are stored inside the device Then tap the Start button to proceed with the measuremen...

Page 57: ... density index Database By tapping the database key you can create a custom database to determine the different density types Tap the New key and enter the name of the new record to be stored Description enter a description of the item Measurement type Select the type of density to be determined After entering all required data tap the Save key to save the new record ...

Page 58: ... Screen with the Density of a liquid calculation function Below are the functions for calculating the density of a liquid Tap the Esc key to exit the function the Restart key to make a new measurement and the Print key to print the measurement result Detected values for density determination Detected values for density determination Information about the database used Record name Test description ...

Page 59: ...nctions for calculating the density index Tap the Esc key to exit the function the Restart key to make a new measurement and the Print key to print the measurement result Detected values for density determination Information about the database used Record name Test description ...

Page 60: ...ghing Database statistics by selecting the statistical mode from the database you can store and then recall a custom statistics method By tapping the database key you can create a custom database to determine the different density types Tap the New key and enter the name of the new record to be stored Description enter a description of the item Nominal weight enter the nominal weight value and sel...

Page 61: ...Man key to enter a manual tare value Tap the T key to run a manual tare Touch the 0 key to reset the plate Tap the End key to end the weighing acquisition and follow the statistical calculation Tap the Esc key to exit the function the Restart key to make a new measurement and the Print key to print the measurement result List of the values captured ...

Page 62: ... the T Man key to enter a manual tare value Tap the T key to run a manual tare Touch the 0 key to reset the plate Tap the End key to end the weighing acquisition and follow the statistical calculation Tap the key to delete the last weight captured Tap the key to display the weights captured Tap the key to display the parameters set for statistical calculation ...

Page 63: ... statistical report After tapping the End key the screen with the result of the weighing statistics is displayed Tap the Esc key to exit the function the Restart key to make a new measurement and the Print key to print the measurement result ...

Page 64: ...ith which you want to view the result Tex Td Nm Nc However the selected unit can still be modified during weighing by tapping the symbol displayed to the left of the weight Start after entering the required data tap the Start button to execute the Textile function Database by touching the database key you can create a custom database for Textile weighing Using the database in addition to the simpl...

Page 65: ...ect the unit of measurement and enter the nominal weight value to be checked T2 T2 T1 T1 enter the error tolerance threshold values Optional entry Save tap the Save button to save the entered record 8 7 1 Screen of the Textile function Below are the functions in the Textile screen Tap the Print key to print the result Tap the T Man key to enter a manual tare value Tap the T key to run a manual tar...

Page 66: ...e a new weight Tap the T Man key to enter a manual tare value Tap the T key to run a manual tare Touch the 0 key to reset the plate Tap the End key to end the weighing acquisition and follow the statistical calculation Tap the key to delete the last weight acquired Tap the key to view the weights acquired Tap the key to display the parameters set for statistical calculation ...

Page 67: ... statistical report After tapping the End key the screen with the result of the weighing statistics is displayed Tap the Esc key to exit the function the Restart key to make a new measurement and the Print key to print the measurement result ...

Page 68: ... weight measurement cycles within the set time interval To access the function from the Functions menu tap the Animal weighing icon Measurement time enter the express measurement time in seconds Minimum value 5 seconds maximum 90 seconds Start after entering the required data tap the Start button to perform the Animal weighing function If necessary reset the indication or load any tare Load the sa...

Page 69: ...p the countdown and display the result Tap the Esc key to exit the function the Restart key to make a new measurement and the Print key to print the measurement result Information during weight capture Countdown indication The average value of the weights captured in the set time interval Information at the end of weight capture Indication of the time used for the measurement ...

Page 70: ...ly displays the measurement screen Prepare the sample to be tested on the weighing plate Perform tare or zero reset if necessary Tap the Ok key to start the test Once the tested sample breakage load is reached the maximum weight will be displayed Weight detected on the plate Tap the Esc key to exit the function the Restart key to make a new measurement and the Print key to print the measurement re...

Page 71: ...u to carry out a series of consecutive weights and finally get the total weight of the dosed components and the relative list with the value of the individual weights Tapping the Simple recipe button immediately displays the measurement screen Load any container and run the tare Or if necessary reset the indication with the 0 key Now dose the first component and tap the Ok key to confirm the captu...

Page 72: ...the component Quantity select the desired unit of measurement and enter the nominal quantity of the component Toll T enter positive percentage tolerance Toll T enter negative percentage tolerance Confirmation tap the confirmation key to accept the entered values Verification of the correctness of the data entered in case of error signal will be performed check that all parameters entered are congr...

Page 73: ... you want to produce 100g insert a K factor of 10 If the entered value does not allow the dosing of one of the components because its quantity exceeds the scale capacity it will not be accepted If the total weight of the recipe instead exceeds the scale capacity but the individual components are lower it will be asked whether to proceed with the separate recipe Yes to proceed with the separate rec...

Page 74: ...o end the weighing capture and follow the statistical calculation Tap the key to clear the last weight acquired Tap the key to view the weight acquired After tapping the End key the screen appears with the result of the sum of the weights Tap the Esc key to exit the function the Restart key to make a new measurement and the Print key to print the measurement result ...

Page 75: ... components already entered Dosing bar The dosing bar works with automatic graduated enlargement When the dosed quantity approaches the two levels of tolerance the center zone is automatically enlarged to allow dosing with a greater sensitivity The bar takes on different colors Yellow when the loaded weight is lower than the minimum threshold Green when the weight is within the set range Red when ...

Page 76: ...confirm the adjustment press the No button to return to the dosing screen In this case you can remove excess quantity or cancel the recipe by tapping the X key When the correct dosage of all components is completed the actual dosed amount will be displayed Tap the key to view the complete list of components and their dosed quantities Tap the Esc key to exit the function the Restart key to make a n...

Page 77: ...wed per parameter is 20 GLP Database by tapping this command you can access the GLP parameter database management menu Below are all the available functions Selection by name y tapping this key you can call up a program by entering the corresponding name Selection list by tapping this button you can call up a program by selecting it from a list New by tapping this button you can enter a new record...

Page 78: ...played Tap the corresponding parameter to enter the desired data Scale ID Project ID User ID After entering the data press the Save key to confirm In order to save data at least one of the three fields must be filled in otherwise an incomplete data screen will be displayed After entering the record to select it use the Select from list or Select by name command Tap the desired record and confirm w...

Page 79: ...ltitude 4000m Degree of Pollution 2 Overvoltage category II Power supply unit supplied INPUT Switching 100 240Vac 50 60Hz OUTPUT 24V DC 550mA Potenza Max absorbed power 13 2VA Adaptation to environmental conditions Selectable filters Autozero Selectable from menu Serial output RS232C Working temperature 5 C 35 C ...

Page 80: ...shocks fall of the balance or fall of objects on the weighing plate The transport to and from the Service Center is the responsibility of the customer 11 Disposal If the packaging is no longer used it can be delivered to the local waste disposal center The packaging is made entirely of non polluting materials recyclable as valuable secondary raw materials Discharged batteries should not be dispose...

Reviews: