MR9610 Microplate Reader Operation Manual Chapter 6 Maintenance, storage, transportation
─
28
─
Chapter 6 Maintenance, storage, transportation
1. Maintenance
⚫
Keep the installation environment dry and clean to prevent moisture, corrosion,
and strong electromagnetic interference sources.
⚫
The SmartReader Instrument is calibrated before it leaves the factory. There are
no user-serviceable parts inside, so please do not attempt to open the housing. If
any defects or operation problems arise, please contact your distributor or Accuris
Instruments at 908-769-5555 or info@accuris-usa.com.
⚫
Do not switch on and off the power abruptly, as this can damage the electronics.
⚫
Make sure to connect the SmartReader to an appropriate power outlet (voltage
and Hz) as stated on the instrument rating plate.
⚫
Maintenance list
Content
Daily
Weekly
Annually
When
needed
Make sure the instrument powers off correctly
√
Keep the instrument away from dust
√
Remove spilled solutions right away to prevent
damage, then clean it with distilled water.
√
If any surface has been contaminated with
biohazardous materials , sterilize it using a
disinfectant.
√
Clean instrument housing regularly.
√
Clean the plate holder when necessary.
√
Maintenance
√
2. Storage and transportation
⚫
When not in use, store the instrument at a temperature range -10
℃
~
40
℃
,
relative humidity less than 80%, without corrosive gas and with good ventilation.
⚫
Prevent from heavy vibrations and humidity during transportation.