59
8.6
“Statistic
s
”
function
This function allows you to perform weighing statistics in simple mode, without defining the
nominal weight and tolerance value, or, using database statistics, you can define the relative
nominal weight and tolerance values. At the end of the procedures in both cases, it is possible
to print the statistical result of the weights.
To access the function from the functions menu, tap the "
Statistics
" icon.
Select the mode:
Simple statistics:
by selecting the simple
function, you can perform the statistics up
to 500 consecutive weights.
Tap the "Simple Statistics" button and
immediately the screen will be activated to
perform the weighing.
Database statistics:
by selecting the
statistical mode from the database, you can
store and then recall a custom statistics
method.
By tapping the "database" key, you can
create a custom database to determine the
different density types.
Tap the "
New
" key and enter the name of
the new record to be stored.
▪
Description:
enter a description of the
item.
▪
Nominal weight:
enter the nominal
weight value and select the desired
unit of measurement.
▪
Enter values for tolerance check
(Optional Data).
▪
T2+ :
2nd positive tolerance
▪
T2- :
1st negative tolerance
▪
T1+ :
1st positive tolerance
▪
T1- :
1st negative tolerance
After entering all required data, tap the
"
Save
" key to save the new record. Before
performing the save, a congruity check of
the entered data is performed. An error will
be reported to make the necessary
corrections and retouch the save key.
T1 -
T2 -
T2 +
T1 +
PESO NOMINALE
Summary of Contents for Tx Series
Page 7: ...6...