SmartLOG Overview

The 

Accutherm SmartLOG 2021

 is a wifi digital data logger (DDL) that monitors and records temperature and sensor battery 

status for up to two wireless sensors. SmartLOG also logs the ambient room temperature/humidity and backup battery voltage. 

The DDL scans for the wireless sensor in 1 minute intervals retrieving temperature, battery and signal data. This data is 

evaluated for, temperature excursions, low battery condition or no data which indicates signal loss from the wireless sensor. 

If any of these conditions are true the SmartLOG will send alerts via email and text message to contacts listed in the settings 

page of the dashboard. The SmartLOG then refreshes the Main DDL display screen and Dashboard with all updated values. 

The SmartLOG Dashboard can be viewed using a PC or Smart Phone web browser. Monitor all DDL conditions, generate PDF 

reports, and modify settings. SmartLOG and PC or Smart Phone must be connected to same local network.

NOTE: If main power is lost, after approximately 4 hours the DDL will turn off the display and wifi radio. This is to 

conserve energy allowing logging to continue until main power resumes. The display can still be viewed with a 

tap, and will show for ~1 minute and then go back off . The Web Dashboard will not be accessible until main power 

resumes 

(connect DDL to a 5v battery bank with usb and this will activate wifi radio alowing Dashboard access

). Alerts will continue to be sent 

while on battery backup. If Temperature Excursion, alarm LED will flash ~10 seconds each min.

The LED for the wireless temperature sensor(s) blinks every 10 seconds indicating the wireless Sensor sent the temperature, 

battery power, and signal strength.

SmartLOG SETUP

Wireless Temperature Sensor(s)

1.  Install quantity 2 AAA batteries by twisting clip cap counterclockwise 1/8 turn to open.

2. 

Place batteries negative end in first.

3.  Align printed arrows on clip cap and sensor to each other, push in and twist 1/8 turn clockwise to close.

4. 

To power on wireless sensor, press power button until blue LED flashes 5 times. LED should flash once every 10 seconds.

 

To power off sensor, press power button until led flashes two times, then release button.

5. 

Note the four digit serial number on the backside of the sensor(s) and note intended location for Fridge, Freezer, other.    

E.g.  Fridge  SN: 0010

6.  Place wireless sensor in storage unit to be monitored. Allow the sensor temperature to stabilize before setting up DDL. 

(wait approx. 60 min)

SmartLOG Digital Data Logger (DDL)

      

NOTE: Following instructions are only necessary on the initial Start-up of SmartLOG All configuration settings are retained in DDL memory.

1.  Install quantity 3 AA batteries in compartment behind DDL. Slide cover down to open. 

Batteries are for backup ONLY! Use wall 110VAC adapter as primary source of power.

2.  Attach wall 110VAC adapter to DDL using USB cable and plug in AC adapter to wall outlet.

3.  Push and hold Power Button on back of SmartLOG DDL until beep sounds… then release.

4.  Screen will show “Scanning For Networks”  

fig. 1

5.  Then will change to “Select Network” when scan completes. 

fig. 2

6.  Tap using provided plastic stylus on “Select Network”

7. 

A drop-down selection of “SSID’s” (wi-fi network names) will appear in location nearest/farthest order. 

fig. 3

8.  Tap on your desired network name.  

9. 

A Password field and Keyboard will appear, enter the wi-fi network password. 

fig. 4

 

Tap eye icon to see typed text if needed.

10.  Tap return key 

 

on keyboard (lower right corner) to enter values. DDL will now attempt to connect to network.

11.  If  SmartLOG is able to make connection, a message will be displayed instructing to complete remaining setup using a 

web browser. 

fig. 5

   Follow “Dashboard using Web-Browser” section to complete settings.

12.  SmartLOG will automatically start scanning for temperature sensors then display Main Screen. 

fig. 7

13.  Tapping “Main” screen anywhere opens the “Options” screen. 

fig. 8

 

For instructions on options go to SmartLOG Options section.

Dashboard using Web-Browser

1.  In Address Bar of a web browser 

fig. 6

 enter “SmartLOG/” or the IP Address displayed on DDL.  

fig. 5

 

Note: Host Name “SmartLOG” can be change in settings page if desired.

2. 

In SmartLOG dashboard, click on “Settings” (cog icon).

3.  Settings page Access Pop-up, leave PIN empty and click “Open”. 

4. 

Complete the form fields, NOTE:  Access Pin field requires a 5 digit PIN number in order to save the form. 

 

This PIN will be required for future access to settings page. 

5. 

Click Save button.  Return to “SmartLOG Digital Data Logger (DDL)” 13 above.

Summary of Contents for SmartLOG 2021

Page 1: ...twork No cloud No subscription fees Enterprise Network Login for Security Time automatically synchronized from internert Reports in PDF format No PC Software needed Second Temperature Sensor optional...

Page 2: ...th 8 Lowest Tempeature Since Last Reset 9 Highest Temperature Since Last Reset 10 Alarm Set Points Min Max 11 Current Temperature P2 Tile Same as P1 1 Dashboard IP Address 2 Close Options Screen 3 Opt...

Page 3: ......

Page 4: ...d flashes two times then release button 5 Note the four digit serial number on the backside of the sensor s and note intended location for Fridge Freezer other E g Fridge SN 0010 6 Place wireless sens...

Page 5: ...being sent out When alerts are complete the screen will switch back to the main screen automatically showing DDL Tile blue and LOGGING OFF replaced wih Logger Name BUZZER Button When set to OFF will...

Page 6: ...RESET MIN MAX Button Sets memory recorded Min and Max values to current tempeature time and date and removes alert and alarm colors from display if excursion was triggered 1 Tap RESET MIN MAX button...

Page 7: ...batteries back in Push on button till you hear beep if no response data logger may be faulty Incorrect time Turn off logger Start logger will update from online time server NOTE DDL DOES NOT adjust fo...

Page 8: ...gging interval WiFi IP addr Mac addr Siginal strength CPU temperature Ram used and free Firmware version running Support Links to online Manual QuickStart video Instructional video REPORT PAGE Month s...

Page 9: ...rd Outgoing SMTP Dropdown list of popular email provider servers Selecting a server name will auto populate the Port Custom server address port can be entered If Alerts and Status and message are not...

Page 10: ...ccount click icon in upper right coner then click View Account On My Account page click Security info On My Sign Ins page click Add method Select App password click Add Enter password Name e g SmartLO...

Page 11: ...rview video Quick Start video Gmail Setup video SmartLOG Kit consist of 1 x SmartLOG Digital Data Logger 1 x Wireless Transmitter 1 x 5V 1A Power Adapter 1 x USB Cable 2 x AAA for transmitter 3 x AA f...

Page 12: ...Direct Sequence Spread spectrum Warranty With respect to the goods unless expressly agreed otherwise in writing Thermco Products Inc provides warranty for material and manufacturing defects for a peri...

Reviews: