eL-720 User’s Guide 

SENSOR 

An 

eL-720

 sensor is designed for a long, trouble-free life.  The 

sensor is tested and calibrated automatically each time the unit is 
powered on.  

A failed or failing sensor will be indicated 3 

seconds after power-on by a two-tone alarm and all LEDs 
flashing. 

 

CLEANING THE SENSOR 

If moisture is suspected of contaminating the sensor, the probe 
and sensor may be cleaned using this procedure. 

1. 

IMPORTANT: 

Remove the sensor from the instrument as 

directed on page 5. 

2.  Remove the probe tip cap and filter from the probe tip. 

3.  Blow compressed air into the probe tip to remove any residual 

moisture from the probe. 

4.  Blow compressed air

*

 into the inlet and outlet ports of the 

sensor. 

5.  Place a few drops of distilled water or isopropyl alcohol into 

both the inlet and outlet ports of the sensor. 

6.  Blow compressed air

*

 into the inlet and outlet ports of the 

sensor to remove any remaining liquid and thoroughly dry the 
exterior of the sensor. 

7.  Install the sensor, and insert a new filter into probe tip. The 

unit should now operate properly. 

 

* Note: Do not exceed 50 psi pressure.  Doing so may result 
in the loss of the small o-rings in the inlet and outlet ports of 
the sensor. 