EC Module
Page 12
Stabilizing …
The unit will then scroll through the flow rate, concentration and gas type,
source life, and battery status.
Flow 0.5 LPM
Select to change
1.00 PPM CL2
Source Life
Battery Status
Note that the gas type will be different depending on the source
used as will the source life and concentration ranges.
At this point, the unit will begin the stabilizing period. The unit will display
the following sequence until either the SELECT button is pressed to
change the flow or concentration, the POWER button is pressed and held
to turn the unit off, or the Standby time elapses.
Stabilizing ...
Standby:60 sec
Flow 0.5 LPM
1.00 PPM CL2
Standby:60 sec
When the Standby time elapses, the display will change to simply display
the concentration as shown below. At this point the calibration gas is
stable and you may proceed with calibration. A timer will run on the
bottom-line indication the time the unit has been running.