background image

Table 4 – User message displayed in Device Screen 

User Action 

Message displayed on 

Device screen 

Description 

REC 

No of REC 

Number of records saved in Device Memory 

SEND 

Records Uploaded 

Total number of records uploaded in Server 

SEND 

Total Records 

Total number of records saved in Device Memory 

SEND 

No Pending Records 

This message will appear if all the records in Device Memory 

have been sent to Server 

 
 

Table 5 – Admin Menu Items and Description 

Sl.No  Menu Item 

Description 

SET TIME&DATE 

To set the current Time and Date  

SET CARD TIMEOUT 

Timeout interval after location ID is captured through RFID scan by User. If 

set to 5 seconds, the Device will display the location ID for 5 seconds soon 

after the Location Tag is scanned. After the timeout, the user must scan 

the Location Tag again for successful recording. 

VIEW IMEI NUMBER  To view device IMEI number 

SET CO2 OFFSET 

To set the offset for CO2 Sensor. To be used only under senior technician's 

presence  

POWER SAVE TIME 

Time out for Backlit to save power while in battery 

VIEW RECORDS 

To view the recorded data present in Device memory 

DELETE RECORDS 

To delete all the records in the device memory. [Separate Password 

protection]   

EXIT 

Exit the Admin Setting screen 

 

Instructions – For Recording 

1.

 

Use the On/Off Switch to turn on the Instrument. Once the Instrument is switched on, 
it displays information as per 

Table 2

 above 

2.

 

Check the Battery status if the device is sufficiently charged. If single bar is blinking 
continuously, please charge the device. [Refer 

Battery Care Section

 for charging 

instruction] 

3.

 

Use the RFID scanner of the device to scan the location ID(RFID) Card. The 
Instrument will display the location ID on screen. 

4.

 

When the location ID is displayed in screen, press the ‘REC’ Button to save the 
current data displaying in the device. The message in screen will display ‘No of REC’ 
i.e. number of records saved in the device memory.  

 

 

Page 2 

Summary of Contents for AI-IAQ-SP-IoT

Page 1: ...ry Bar Indicator 4 Bars 4 Bars Full Battery 1 Bar blinking Low Battery 5 Power Indicator Indicates as P when connected to power 6 Temperature Indicator Indicates in Degree Celsius 7 Humidity Indicator Indicates in RH Relative Humidity 8 CO2 Indicator Indicates in PPM parts per million Table 3 Basic operations of Front keys Sl No Button Function 1 Enter 1 Enter in Admin Setting Screen 2 Saving Valu...

Page 2: ...umber 4 SET CO2 OFFSET To set the offset for CO2 Sensor To be used only under senior technician s presence 5 POWER SAVE TIME Time out for Backlit to save power while in battery 6 VIEW RECORDS To view the recorded data present in Device memory 7 DELETE RECORDS To delete all the records in the device memory Separate Password protection 8 EXIT Exit the Admin Setting screen Instructions For Recording ...

Page 3: ...e Instructions For posting Data to Server 1 Ensure sufficient Internet data balance is available in the SIM card before using the Instrument 2 Insert GSM sim card in the SIM slot make sure the SIM is of standard size i e 15 X 25 mm Ensure SIM s corner cut is facing the charger slot Refer Pg 6 for inserting 3 Plug the charging cable to device and check if network signal is displayed in the Instrume...

Page 4: ...s is strictly recommended for avoiding any damage to battery 2 Charge only if the battery shows 1 blinking bar or if the Instrument is not switching ON RFID Care 1 Do not expose RFIDs in higher temperature than 50 Deg C 2 Do not bend the RFID tags Basic Troubleshooting 1 Sensor not responding Blow air gently in both the sensors placed at the top of Instrument and check if values are changing 2 Net...

Page 5: ...Sagar Sharma Technical Head 9989190588 sagar instrukart com within 6 hours 3 L3 Chetan Sachdeva Managing Partner 9394030848 chetan instrukart com within 12 hours Return Address of Instrument In case of any repair Re calibration after 12 months INSTRUKART HOLDINGS Plot no 18 Road no 1 A Czech colony Near Reliance Fresh Sanath Nagar Hyderabad 500018 Phone 040 40262020 7331110506 Note Please mention ...

Page 6: ...comprehensive Instruction Manual video please follow https www youtube com watch v L1JfFXBZV2g CO2 Sensor Temp Humidity Sensor Power Slider Charging In SIM Slot Enter Up Right Exit REC SEND USB Out Antenna Insert this way ...

Reviews: