Note:

 

 

 

The location ID will display for 7 seconds only. Ensure you press the ‘REC’ 
button within this time period else the location ID needs to be scanned again 
through the RFID 

 

Pressing the Rec Button multiple times in quick succession will save multiple 
data from the same location.  

 

Pressing ‘Rec’ button when Location ID is not displayed in the device will not 
save the intended data i.e. along with location ID. 

 

User needs to wait for at least 2 minutes before any reading to be taken at 
any location i.e. AC Location, Ambient Location, Moving between AC to 
Ambient, AC location to Ambient.  

5.

 

Carry the device to another location/ODC  to capture the next record. Repeat 
steps 3 & 4 to perform the operation again. 

Important Note:

 Please ensure the Instrument’s Sensors (Temp, RH & CO2) are kept away 

from sources of CO2 (Mouth, Nose etc..) as the actual result/recording data will be 
changed. The CO2 sensors are highly sensitive in its properties and it is always 
recommended that the Instrument is kept away from direct and close distance exposure 
to Mouth and Nose. 

Instructions – For posting Data to Server 

1.

 

Ensure sufficient Internet data balance is available in the SIM card before using the 
Instrument.  

2.

 

Insert GSM sim card in the SIM slot(make sure the SIM is of standard size i.e. 15 X 25 
mm). Ensure SIM’s corner cut is facing the charger slot. Refer Pg 6 for inserting. 

3.

 

Plug the charging cable to device and check if network signal is displayed in the 
Instrument . If not, please check the network or SIM.  

Note: The availability of network in the device is subject to transmission from Carrier 
(Airtel, Vodafone, Idea etc..) and the strength in the region  

While the data is being uploaded, do not remove the charging cable or press any 
button.  

4.

 

On visibility of the network signal in the device (minimum of 4 bar), press the ‘SEND’ 
button for posting recorded data to Server. The screen will display ”Uploading 
Records”. 

5.

 

During the upload process, the Device may stop the upload and screen remains 
silent. This could happen because of network connectivity with mobile 
operators(Airtel, Vodafone etc). In this case, please restart the device using 
ON/OFF switch and the Instrument will resume the upload process. If the uploading 
process is not resumed, press SEND button once and wait for sometime for the 
operation to complete. 

Summary of Contents for AI-IAQ-SP-IoT

Page 1: ...ry Bar Indicator 4 Bars 4 Bars Full Battery 1 Bar blinking Low Battery 5 Power Indicator Indicates as P when connected to power 6 Temperature Indicator Indicates in Degree Celsius 7 Humidity Indicator Indicates in RH Relative Humidity 8 CO2 Indicator Indicates in PPM parts per million Table 3 Basic operations of Front keys Sl No Button Function 1 Enter 1 Enter in Admin Setting Screen 2 Saving Valu...

Page 2: ...umber 4 SET CO2 OFFSET To set the offset for CO2 Sensor To be used only under senior technician s presence 5 POWER SAVE TIME Time out for Backlit to save power while in battery 6 VIEW RECORDS To view the recorded data present in Device memory 7 DELETE RECORDS To delete all the records in the device memory Separate Password protection 8 EXIT Exit the Admin Setting screen Instructions For Recording ...

Page 3: ...e Instructions For posting Data to Server 1 Ensure sufficient Internet data balance is available in the SIM card before using the Instrument 2 Insert GSM sim card in the SIM slot make sure the SIM is of standard size i e 15 X 25 mm Ensure SIM s corner cut is facing the charger slot Refer Pg 6 for inserting 3 Plug the charging cable to device and check if network signal is displayed in the Instrume...

Page 4: ...s is strictly recommended for avoiding any damage to battery 2 Charge only if the battery shows 1 blinking bar or if the Instrument is not switching ON RFID Care 1 Do not expose RFIDs in higher temperature than 50 Deg C 2 Do not bend the RFID tags Basic Troubleshooting 1 Sensor not responding Blow air gently in both the sensors placed at the top of Instrument and check if values are changing 2 Net...

Page 5: ...Sagar Sharma Technical Head 9989190588 sagar instrukart com within 6 hours 3 L3 Chetan Sachdeva Managing Partner 9394030848 chetan instrukart com within 12 hours Return Address of Instrument In case of any repair Re calibration after 12 months INSTRUKART HOLDINGS Plot no 18 Road no 1 A Czech colony Near Reliance Fresh Sanath Nagar Hyderabad 500018 Phone 040 40262020 7331110506 Note Please mention ...

Page 6: ...comprehensive Instruction Manual video please follow https www youtube com watch v L1JfFXBZV2g CO2 Sensor Temp Humidity Sensor Power Slider Charging In SIM Slot Enter Up Right Exit REC SEND USB Out Antenna Insert this way ...

Reviews: