6.8 Resolution Switch
Action
To switch scale’s resolution, press
and
on remote controller in turn.
Function
The screen flashes the resolution to change in descending order, like 5kg, 2kg, 1kg (available optional resolution
ranges from 0.01 kg or lb to 50 kg or lb, depending on scale’s capacity). After Resolution Switch, new resolution
takes effect.
To balance between accuracy and measurement speed in high resolution mode, Anti-Motion Level can be
changed to fit your measurement application.
To learn how to set Anti-Motion Level, please refer to section 10.6 – Anti-Motion Level.
Condition
The scale must not be in Hold mode. Otherwise, error message
will keep flashing.
High resolution offers better accuracy at the cost of
longer measuring time and stricter requirement of
load’s stability. Designed to meet the OIML R76’s
directive, this cale has the best (default) performance at
the 2000 to 3000 division.
RESOLUTION SWITCHING
will change the apparent
overloading range of the scale. For example, if the resolution
of a 3000kg scale is switched to 0.5kg, it will overload at a
new weight of 3004.5kg (3000 + 9 * 0.5), while by default, it
overloads at a weight of 3009kg (3000 + 9 * 1).