OMNISWD
®
USER MANUAL
23
COPYRIGHT © 2019-2020, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Upon completion of tuning, user is prompted to select DONE. When DONE is pressed, User is returned to the
previous screen viewed before Warning was generated. User is brought back to previous screen they were in before
Caution pop-up warning was prompted.
Output Tuning is Required (>20% variance)
If the Output tuning is greater than >20%. An “Output tuning is
required” will pop-up on the screen, It is required to tune the
device to continue treatment.
Please select TUNE to be taken to the Device Output Tuning
screen.
Press the Start button below. Gradually turn the Tuning Knob
on the applicator head to display the most amount of green and
the least amount of yellow on the bar below. Adjust the tuning
dial for minimum reflected power, while achieving forward
power as close to 200 watts as possible. The goal for reflected power is less than or equal to 2 watts, but anything
less than 10 watts is acceptable. Press DONE when complete.
Upon completion of tuning, user is prompted to select DONE. When DONE is pressed, User is returned to the
previous screen viewed before Caution pop –up warning was generated.
End Of Treatment: QR Code
Systems Information
A screen with the information will appear. The System
information screen includes Serial Number, Trade
Name / Model Number, Bootloader Version, Display
version, Systems version, WiFi Module Version, Last
calibration, utilization Tier, ACP Customer Support,
and the ACP Website address.
At the end of treatment a QR code Pop-can be scanned
to record the treatment into the Electronic Medical
Record via the ACPlus iOS application.