26 -
OMNIVERSA
®
ULTRASOUND / ELECTROTHERAPY SYSTEM USER MANUAL
COPYRIGHT © 2017-2020, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
Pausing a Treatment
As soon as the Start
button is pressed the PAUSE
and STOP
button appear. To pause a treatment, touch the
PAUSE
button.
A two bar orange PAUSE
Icon will
appear at the top of in the channel tab.
When PAUSE
is selected output intensity ramps
down to 0 and the START
and STOP
buttons reappear.
To resume the paused treatment, press the START
button, touch the channel tab(s) to illuminate orange and reset
the output intensity.
After pressing PAUSE
do not remove electrodes for
3 seconds to allow any remaining capacitance to dissipate.
Note:
To end the treatment, touch the STOP
button. A red
square STOP
Icon will appear at the top of the channel
tab.
The treatment can also be stopped by pressing the emergency
stop icon
.
Pressing this button stops the output on all
channels in use.
Turn Off Unit
Do not turn off the unit by means of the mains switch, as it will
continue operation from the battery power. The current will just keep
flowing through the patient as if nothing happened and thus it will also
flow through the therapist, causing them to inadvertently feel the
current if they attempt to remove electrodes.
It is recommended that user turn off the unit by either one of three
ways: Turning off the unit by: 1) pressing the orange off button on the
top rear of the unit, 2) pressing the stop symbol
on the upper
right hand side of the display, or 3) by pressing the Stop button on the
lower right hand side of the display:
All three ways will cause the unit to immediately switch off the output
power and drain any remaining capacitance. Doing this will therefore
make it impossible for the therapist to feel a shock from remaining
output power.