OMNIVERSA
®
ULTRASOUND / ELECTROTHERAPY SYSTEM - 45 -
COPYRIGHT © 2017-2020, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
channels. Output stimulation from Ch. A+B occurs
Simultaneously, but Ch. B has a Delayed start.
Channel Interactions – Single Channel Operation
When a single channel setup (Ch. A or B only) are selected Channel
Interactions are as follows.
Manual:
Ch. A Ramp Up and Ramp Down Time are preset to
2-seconds. Stimulation On time (Start Estim) and Off time
(Stop Estim) are controlled by the clinician using either the
Clinician Hand Control or commands on the Main Display.
Sequential Simultaneous
: Ch. A Ramp Up and Ramp Down
Time preset to 2-seconds. Stimulation On Time (Start Estim)
is preset to 10-seconds. Stimulation off time controlled by the
clinician using either the Clinician Hand Control or
commands on the Main Touchscreen Display.