ULTRASOUND / ELECTROTHERAPY SYSTEM - 51 -
COPYRIGHT © 2017-2020, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
ELECTRODES ARE FOR INDIVIDUAL PATIENT USE ONLY!
Do not turn power on, or power-off the Stimulator while electrodes are applied and the patient is connected to
the unit.
Use a quality, high conductivity electrode specified by the manufacturer (ACP).
If decreased sensation is suspected, test prior to application of stimulation. Treatment areas with absent or
diminished sensation should be avoided or treated with caution.
For maximum comfort, use the largest electrodes possible without causing overflow to non-targeted areas.
Place electrodes a minimum of 1” apart. Extend the distance between electrodes to achieve deeper current flow
and/or stimulation of larger areas or multiple muscles. Do not place electrodes under hot or cold packs.
Do not use small electrodes (less than 2”x2”) when treating with medium frequency currents (MFAC or IFC).
•
Patient may perform therapist-directed exercises in a relaxed toe down position and/or functional
activities during PENS induced muscle contractions
•
TO ACCOMPLISH A TOE DOWN POSITION WHILE SEATED: PLACE A TOWEL ROLL UNDER HEEL,
PLACE A PILLOW UNDER THIGH, RAISE SEAT HEIGHT
ANKLE DORSIFLEXION
ANKLE DORSIFLEXION
AND EVERSION
PLANTAR FLEXION
ANTERIOR
AND
INFERIOR
TO THE
FIBULAR
HEAD
MID
TIBIALIS
ANTERIOR
ANTERIOR AND
INFERIOR TO
THE
FIBULAR HEAD
MID TIBIALIS
ANTERIOR
PERONEUS
LONGUS &
BREVIS
CH A
CH B
CH A
TRICEPS SURAE
(GASTROCNEMIUS
SOLEUS)