OMNIVERSA
®
ULTRASOUND / ELECTROTHERAPY SYSTEM - 7 -
COPYRIGHT © 2017-2020, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
ELECTROTHERAPY INDICATIONS & CONTRAINDICATIONS
CAUTION: Federal law restricts this device for sale or use by or on the order of a Practitioner
licensed by the laws of the state in which he/she practices to use or order the use of the device.
Please notice that Accelerated Care Plus cannot provide medical advice. If you have specific
medical questions, please contact your healthcare professional.
Indications
The OmniVersa is indicated for:
Relaxation of muscle spasms
Re-education of muscle
Prevention or retardation of disuse atrophy
Immediate post-surgical stimulation of calf muscles to prevent venous thrombosis
Increase local circulation
Maintaining or increasing range of motion
Symptomatic relief and management of chronic intractable pain and as an adjunctive treatment in the
management of acute pain, post-surgical pain and pain associated with post-traumatic injury.
Electrical muscle stimulator devices should only be used under medical supervision for adjunctive therapy for the
treatment of medical diseases and conditions.
Contraindications
Do not use this device on patients who have a cardiac pacemaker, implanted defibrillator, or other implanted
metallic or electronic device because this may cause electric shock, burns, electrical interference, or death.
Note:
There is no contraindication to the application of Transcutaneous Electrical Stimulation or Powered Muscle
Stimulation over metal implants.
Never connect lead wires to the power line or high frequency electro-surgery equipment. Simultaneous
connection of a patient to a high frequency surgical or medical equipment and the stimulator may result in
burns and possible damage to the stimulator. Use only the lead wires recommended or approved by the
manufacturer
Adverse Reactions
Skin irritation and burns, beneath the electrodes, have been reported with the use of powered muscle
stimulators. Patients may experience skin irritation and burns beneath the stimulation electrodes applied to
the skin.
Patients may experience headache and other painful sensations during or following the application of
electrical stimulation near the eyes and to the head and face.
Patients should stop using the device and should consult with their physicians if they experience adverse
reactions from the device.