OMNIVERSA
®
MULTI-MODALITY THERAPY USER MANUAL
- 81 -
COPYRIGHT © 2017-2020, ACCELERATED CARE PLUS CORP., ALL RIGHTS RESERVED
INCREASE LOCAL CIRCULATION
HVPC Sensory
Increases local circulation by stimulating vasodilation. Application of negative polarity over the edematous site
has been demonstrated to reduce post-traumatic edema in clinical trials. In addition, if pain is present, can reduce
it via segmental pain modulation mechanisms. Set amplitude to as high as comfortably tolerated. Increase
amplitude during the treatment in response to accommodation to the electrical stimulation, when and as needed.
Bipolar either side of trauma or black (negative) electrode over edematous tissue, red (positive) adjacent.
Adjustable
Parameter
(Y/N)
Parameter Name
Range (unit)
Default (unit)
Y
PULSE RATE
1…200 (Hz)
125 (Hz)
Y
TREATMENT TIME
0…60 (min)
45 (min)
PENS Muscle Pump
This protocol provides gentle high repetition muscle contractions of stimulated muscle thereby increasing local
circulation. Apply channels and lead wire polarities to the electrodes on the largest muscle groups in the targeted
body area, Channel A on the agonist and Channel B on the antagonist. Set for moderate comfortable muscle
contractions. Adjust amplitude during the treatment in response to the electrical stimulation, when and as needed.
MFAC Muscle Pump
Increases venous and lymphatic circulation via electrical stimulation-induced muscle contractions, which act as
a ‘pump’ and compress fluids into the venous and lymphatic systems. The “Off Time” default is 4 secs. Change
the “On Time” from 4 secs to 12 secs with 2 sec ramp-up and ramp-down times by selecting Channel Timing
and highlighting the “On time”, "ramp-up time" and "ramp-down time" icons and turning the rotatory dial to 12
sec, 2 sec and 2sec respectively. Moderate muscle contraction. Increase amplitude during the treatment in
response to accommodation to the electrical stimulation, when and as needed. Bipolar over muscles in target
tissue area. If using 2 Channels, treat the agonist and the antagonist.
Adjustable
Parameter
(Y/N)
Parameter Name
Range (unit)
Default
units
Y
PHASE DURATION
40…100 (μs)
70 (μs)
Y
RPM
10-120 rpm
30 (rpm)
N
PULSE RATE
50 (Hz)
Y
TREATMENT TIME
0…60 (min)
15(min)
Adjustable
Parameter
(Y/N)
Parameter Name
Range (unit)
Default
(unit)
N
CARRIER FREQ
5.0 (kHz)
Y
BURST FREQUENCY
0.2…200 (Hz)
35 (Hz)
Y
TREATMENT TIME
0…60 (min)
15 (min)
Y
CHANNEL MODE TIMING:
(S)Simultaneous, (A)Alternating
(D)Delayed,
S,A,D
A
Y
B OUTPUT DELAY
0…69 (s)
0 (s)
Y
RAMP UP
0…9 (s)
1 (s)
Y
RAMP DOWN
0…9 (s)
1 (s)
Y
T-ON
0…60 (s)
4(s)
Y
T-OFF
0…120 (s)
4 (s)